32
CHAPTER 6. FAULT DETECTION
6.1.2 Detector Over Range
If the detector is exposed to high levels of radiation or has an internal malfunction that
causes it to count beyond its count rate ability or dead time correction causes a math er-
ror, the unit flashes "or." The user should ensure whether a high radiation filed, internal
malfunction, or invalid settings are causing this fault by observing the instrument’s behavior
in a known radiation field.
Figure 6.2: Fault - Over Range.
6.1.3 Detector Overload
If a current overload threshold is enabled, the instrument can detect when there are some
forms of detector failure, a cable failure, or a high level of radiation. In this event, the unit
will flash "oL."
Figure 6.3: Fault - Overload.
6.1.4 Calibration Expiration
If the calibration expiration lockout is enabled, the instrument will flash "CAL" when cali-
bration is past due.
ò
DIP switch must be enabled. See Chapter 8, Calibration Lockout on page 56 for
more information.
Ludlum Measurements, Inc.
Summary of Contents for 48-4203
Page 2: ......
Page 4: ......
Page 6: ......
Page 10: ......
Page 12: ......
Page 14: ......
Page 19: ...C H A P T E R 3 QUICK REFERENCE 3 1 Instrument Front Diagram 15 ...
Page 20: ...16 CHAPTER 3 QUICK REFERENCE 3 2 Instrument Rear and Sides Diagram Ludlum Measurements Inc ...
Page 21: ...3 3 INSTRUMENT INTERNAL DIAGRAM 17 3 3 Instrument Internal Diagram June 2022 ...
Page 22: ......
Page 26: ......
Page 34: ......
Page 38: ......
Page 58: ......
Page 64: ......
Page 67: ...9 1 INSTRUMENT 63 June 2022 ...
Page 70: ......
Page 76: ......
Page 88: ......
Page 90: ......
Page 91: ......
Page 93: ......
Page 95: ......
Page 97: ......
Page 99: ......
Page 100: ......