4.6. SETUP PAGE 1
39
ð
Parameters on Page 5 are not user editable using the instrument Setup Mode.
To edit these parameters, please use the Ludlum Lumic Calibration or Lumic
Datalogging Software.
4.6 Setup Page 1
Calibration Constant Mantissa -
A short press of ON/ACK to select the value to adjust, and
MODE to adjust the value. Value is used with Calibration Constant Exponent to express
counts per R. Available values are:
• Ones Place (0-9)
• Tens Place (0-9)
• Hundreds Place (0-9)
Calibration Constant Exponent -
A short press of ON/ACK to select the value to adjust,
and MODE to adjust the value. Value is used with Calibration Constant Mantissa to express
counts per R. Available values are:
• Ones Place (0-9)
• Tens Place (0-9)
Dead Time Correction 1 -
A short press of ON/ACK to select the value to adjust, and MODE
to adjust the value. Value is in microseconds. Value should be set to zero when checking
instrument with a pulser or function generator; otherwise, higher count rates will result in
non-linearities. Available values are:
• Ones Place (0-9)
• Tens Place (0-9)
• Hundreds Place (0-9)
See the following table for examples with specific detectors.
Detector
Dead Time
Calibration Constant
Model 44-2
10
µ
sec ±5
µ
sec
100e+08
Model 44-6
100
µ
sec ±35
µ
sec
709e+05
Model 44-7
262
µ
sec ±35
µ
sec
139e+06
Model 44-9
85
µ
sec ±35
µ
sec
200e+06
Model 44-38
100
µ
sec ±35
µ
sec
709e+05
Table 4.2: Dead Time and Calibration Constant
December 2021
Summary of Contents for 25009185
Page 2: ......
Page 4: ......
Page 6: ......
Page 7: ......
Page 8: ......
Page 12: ......
Page 14: ......
Page 32: ......
Page 35: ...29 Weight 0 7 kg 1 6 lb excluding detectors and clips December 2021 ...
Page 36: ......
Page 56: ......
Page 62: ......
Page 72: ......
Page 74: ......
Page 78: ......