Adjustment and Maintenance
4
4.5.1.DROPPER CONTROL MODES
4.5.1.1.STANDBY
4.5.1.1.1.To Select
This mode is chosen upon power-up, when the front panel RESET button is
pressed, or when a time-out has occurred (usually indicating failure).
4.5.1.1.2.Function
The controller is in standby. The motor is turned off.
4.5.1.1.3.To Deselect
Press the front panel INIT button.
(Note: The initialize function can also be executed remotely through the INIT
BNC connector.)
!
WARNING: IT IS DANGEROUS TO ALLOW THE COMPUTER
REPEATED INITIALIZE CAPABILITY. THE SYSTEM CAN ENTER THIS
STATE THROUGH TIME-OUT, WHICH MAY INDICATE SYSTEM FAILURE.
4.5.1.2.MANUAL
4.5.1.2.1.To Select:
THIS MODE SHOULD NOT NORMALLY BE NEEDED BY THE USER AND
SHOULD NOT BE USED WITHOUT FULL UNDERSTANDING OF THE
CONTROLLER!
This mode is selected by turning the front panel selector switch to MANUAL and
pressing the INIT button.
4.5.1.2.2.Function:
The front panel potentiometer controls a servo position for the cart. The cart
servo will not initiate until the trimpot is within a predefined window of the
actual cart position. This will time out after 20-30 seconds if the time out is
enabled (default).
IF THE TIME OUT IS DISABLED, BE CAREFUL TO NOT BURN OUT THE
MOTOR BY DRIVING IT INTO THE STOPS AT THE TOP OR BOTTOM.
4.5.1.3.DROP
4.5.1.3.1.To Select:
Set the selector to AUTO. Press the INIT button to initiate the drop mode.
4-27
Summary of Contents for Micro-g LaCoste FG5
Page 7: ...Table of Figures This Page is Intentionally Blank vii ...
Page 8: ......
Page 13: ...Design Components and Function 2 Figure 2 3 Side view of the dropping chamber 2 3 ...
Page 23: ...Design Components and Function 2 Figure 2 9 The Superspring 2 13 ...
Page 31: ...Design Components and Function 2 Figure 2 12 Rotation Monitor 2 21 ...
Page 32: ...Design Components and Function 2 This Page is Intentionally Blank 2 22 ...
Page 36: ...How to Set Up and Run the FG5 3 Figure 3 1 Superspring Interferometer Setup 3 4 ...
Page 56: ......
Page 59: ...Adjustment and Maintenance 4 4 3 ...
Page 78: ...Adjustment and Maintenance 4 4 22 ...
Page 106: ......
Page 138: ...Checklists and Logs Appendix D Measure gap between two levers near ZPM 9 24 ...
Page 143: ...Checklists and Logs Appendix D This Page is Intentionally Blank 9 29 ...
Page 144: ......