Remove motor mount assembly (motor, Helical coupling, travel
lock plate, and shaft encoder) from Conflat flange.
Loosen screw on Helical coupling (where it attaches to drive pulley
shaft).
Remove Conflat mounting screws.
Remove Conflat (with feedthrough attached).
Remove Helical coupling from feedthrough.
Unscrew feedthrough from Conflat flange.
Lubricate O-ring on new feedthrough. Gently tighten new
feedthrough to Conflat flange (use pliers with padded jaws).
Table 9-8 Replacing Ferrofluidic Feedthrough
9.4.6.Replacing V-plate
Replacing V-plate Checklist
Remove dropping chamber cover.
Remove bottom drag-free cover from cart.
Detach LED bracket from cart and pull out of the way (be careful
not to damage wires).
Observe position of the threaded post attaching top drag-free cover
to cart.
9-10
Summary of Contents for Micro-g LaCoste FG5
Page 7: ...Table of Figures This Page is Intentionally Blank vii ...
Page 8: ......
Page 13: ...Design Components and Function 2 Figure 2 3 Side view of the dropping chamber 2 3 ...
Page 23: ...Design Components and Function 2 Figure 2 9 The Superspring 2 13 ...
Page 31: ...Design Components and Function 2 Figure 2 12 Rotation Monitor 2 21 ...
Page 32: ...Design Components and Function 2 This Page is Intentionally Blank 2 22 ...
Page 36: ...How to Set Up and Run the FG5 3 Figure 3 1 Superspring Interferometer Setup 3 4 ...
Page 56: ......
Page 59: ...Adjustment and Maintenance 4 4 3 ...
Page 78: ...Adjustment and Maintenance 4 4 22 ...
Page 106: ......
Page 138: ...Checklists and Logs Appendix D Measure gap between two levers near ZPM 9 24 ...
Page 143: ...Checklists and Logs Appendix D This Page is Intentionally Blank 9 29 ...
Page 144: ......