background image

Rel. 20220215

Rekalibrierungs-Anleitung

Instruc�ons for recalibra�on

für Temperaturkalibratoren

for temperature calibrators

DEUTSCH
ENGLISH

Seite 2 ff.
page 6 ff.

 

FLUID

LR-

 

Cal

 

QUARTZ

LR-

 

Cal

 

PULSAR

LR-

Cal 

 

SOLAR

LR-

 

Cal

jeweils alle Modelle ab Januar 2021
all models from January 2021 

DRUCK TEMPERATUR

 & 

 Leitenberger GmbH • 

Bahnhofstr. 33 • D-72138 Kirchentellinsfurt • GERMANY

Tel. +49 (0) 7121-90920-0 • Fax +49 (0) 7121-90920-99 •  [email protected] • 

DT

www.druck-temperatur.de

MANUAL

Kalibrierung von Temperaturkalibratoren
Calibra�on of temperature calibrators

1

Summary of Contents for FLUID 100-N

Page 1: ...UARTZ LR Cal PULSAR LR Cal SOLAR LR Cal jeweils alle Modelle ab Januar 2021 all models from January 2021 DRUCK TEMPERATUR Leitenberger GmbH Bahnhofstr 33 D 72138 Kirchentellinsfurt GERMANY Tel 49 0 71...

Page 2: ...r Wenn Sie sich in der 2 Men Ebene be nden siehe Anleitung zu Ihrem Kalibrator dr cken Sie gleichzei g die Tasten und F sobald Sie den KEY Parameter erreicht haben Dr cken Sie die Taste um durch das M...

Page 3: ...play angezeigt 3 1 Kalibrierung des internen Temperatursensors F hren Sie den F hler des Pr zisions Referenzthermometers in die nung des Kalibrierbads oder in eine passende nung des Metallblocks ein S...

Page 4: ...Taste E Wechseln Sie in die 3 Men Ebene und Sie mit den Tasten stellen oder den auf Best gen Sie durch Dr cken der Taste Cal_chnl EXT E Legen Sie zwei Kalibriertemperaturen in Abh ngigkeit Ihres Arbei...

Page 5: ...ein Legen Sie zwei Temperaturpunkte fest abh ngig von Ihrem Arbeitsbereich oder dem Temperaturbereich Ihres Kalibrators Stellen Sie in der 1 Men Ebene die Solltemperatur Setpoint auf den ersten Temper...

Page 6: ...ply only to the following temperature calibrators LR Cal FLUID all models and versions LR Cal QUARTZ all models and versions LR Cal PULSAR all models and versions LR Cal SOLAR all models and versions...

Page 7: ...be shown in the display 3 1 Calibra on of the internal temperature sensor Insert the probe of the precision reference thermometer into the opening of the calibra on bath or into a suitable opening of...

Page 8: ...ing the key E Change to the 3rd menu level and use the or keys to set the to Con rm by pressing the key Cal_chnl EXT E De ne two calibra on temperatures depending on your working range or the temperat...

Page 9: ...ening or into a suitable hole in the metal block De ne two temperature points depending on your working range or the temperature range of your calibrator In the 1st menu level set the target temperatu...

Page 10: ...hofstr 33 D 72138 Kirchentellinsfurt GERMANY Tel 49 0 7121 90920 0 Fax 49 0 7121 90920 99 Info Leitenberger de DT www druck temperatur de MANUAL Kalibrierung von Temperaturkalibratoren Calibra on of t...

Page 11: ...hofstr 33 D 72138 Kirchentellinsfurt GERMANY Tel 49 0 7121 90920 0 Fax 49 0 7121 90920 99 Info Leitenberger de DT www druck temperatur de MANUAL Kalibrierung von Temperaturkalibratoren Calibra on of t...

Page 12: ...NUAL Kalibrierung von Temperaturkalibratoren Calibra on of temperature calibrators 12 DRUCK TEMPERATUR Leitenberger GmbH Bahnhofstr 33 D 72138 Kirchentellinsfurt GERMANY Tel 49 0 7121 90920 0 Fax 49 0...