Operation

EN

5.1 Patient menu 

1. To call up the Patient menu, press the menu key 

briefly

. Total therapy time is displayed.

2. To scroll through the Patient menu and switch to the next parameter, press the Menu key 

again.

3. To exit the Patient menu, press the On/Off/ESC key.

 

Alternatively,

 

if you press no key for 30 seconds, the Patient menu closes automatically. 

Parameter

Display

Value range

Key command

(alternative)

Described in

Leakage

Le

Low, moderate, 
high

Press menu key briefly 
with therapy device 
switched on

Section 5.7, on 
page 25

Therapy time

h

Section 5.2, on 
page 21

Autostart

Auto

0n
0FF

Section 5.3, on 
page 21

Mask test

test

--, 8, 12, 16, 
20 hPa

Section 5.4, on 
page 22

Softstart time

min

 

5 to 45

With therapy device 
switched on: press 
and hold 

Section 5.5, on 
page 23

Humidifier stage

1 to 6

With therapy device 
switched on

: press 

and hold 

Section 5.6, on 
page 23

Drying mode

dr 0:30

With therapy device 
switched off: 

 

press 

 and 

 at 

the same time

Section 6.2, on 
page 30

Summary of Contents for SOMNOclick 300

Page 1: ...ber 65 000 SOMNOvent auto S ST automatic BiLevel S ST therapy device with autoTriLevel principle SOMNOvent auto S SOMNOvent auto S with SOMNOclick 300 SOMNOvent auto S 230 V SOMNOvent auto ST SOMNOven...

Page 2: ...t 22 5 5 Softstart setting 23 5 6 Respiratory air humidifier 23 5 7 Operate therapy device 25 5 8 After use 27 5 9 Travel with the therapy device 28 6 Hygiene treatment 29 6 1 Intervals 29 6 2 Breathi...

Page 3: ...port 14 Head strap 13 Breathing mask 9 Click adapter 1 Bacteria filter 4 Control panel and displays 5 Connection for respiratory air humidifier 8 Drying adapter 7 Sealing plug 10 Pressure measure ment...

Page 4: ...nnection for respiratory air humidifier For connecting SOMNOclick 300 to the therapy device 6 Device outlet port with pressure measurement connection The respiratory air flows to the patient from here...

Page 5: ...midifier This symbol appears when the humidifier is in op eration The set humidifier stage is displayed 19 On Off key For switching the therapy device on and off and for exiting a menu 20 Menu key For...

Page 6: ...ort outlet for ambient air at 4 20 hPa 33 Connection for adjusting therapy parameters using WEINMANNadjust PC software or SOMNOadjust remote setting for professional staff to read out the course of th...

Page 7: ...te Contains useful tips Symbol Meaning Year of manufacture Degree of protection against electric shock device type B Do not dispose of the device in domestic waste Follow the relevant instructions for...

Page 8: ...ng ventilation Note To ensure successful therapy with SOMNOvent auto S it is important to determine and set accordingly therapy mode pressure limits and minimum respiratory frequency SOMNOvent auto ST...

Page 9: ...s different modes CPAP S one of three pressure variants can be selected for S mode BiLevel IPAP and EPAP can be set to a fixed value TriLevel IPAP EPAP and EEPAP can be set to a fixed value auto TriLe...

Page 10: ...e or reduce trigger sensitivity A different speed can be set for pressure rise on inspiration if required Auto On Off can be activated on the therapy device The therapy device can be switched on by a...

Page 11: ...e connected to an easily accessible socket so that the plug can be disconnected quickly in the event of a fault Do not set up the therapy device close to a radiator and do not expose it to direct sunl...

Page 12: ...midifier may run into the therapy device and damage it if it is tilted Accessories replacement parts Caution If third party items are used functional failures and restricted fitness for use may result...

Page 13: ...octor supervising treatment to decide whether use is indicated In such cas es use needs to be tightly controlled and the risk weighed up against the benefits This implies Acute cardiac decompensation...

Page 14: ...sal cannula and forehead cushion reddening of facial skin blocked nose dry nose dry mouth in the morning sensation of pressure in the sinuses irritation of the conjunctiva of the eyes gastrointestinal...

Page 15: ...pply 1 Use the power cord to connect the power supply con nection of the therapy device to a socket The therapy device automatically detects whether 115 V or 230 V are connected and automatically swit...

Page 16: ...ry air humidifier Note Follow the instructions for filling and adapting the SOMNOclick 300 respiratory air humidifier in the associated instructions for use 4 4 Breathing mask Put on breathing mask 1...

Page 17: ...n system the CO2 concentration in the mask and tube may rise to critical values and thus obstruct your breathing Risk of asphyxiation Fit the external exhalation system between the breathing mask and...

Page 18: ...alve WM 24042 is used up to 4 l min oxygen can be supplied to the breath ing mask It is essential to follow the safety instructions for handling oxygen as well as the instructions for use for the oxyg...

Page 19: ...ly by a key command as an alternative to going through the Patient menu Patient menu in standby mode 5 2 Read out therapy time on page 21 5 3 Auto On Off on page 21 5 4 Mask test on page 22 5 5 Softst...

Page 20: ...ange Key command alternative Described in Leakage Le Low moderate high Press menu key briefly with therapy device switched on Section 5 7 on page 25 Therapy time h Section 5 2 on page 21 Autostart Aut...

Page 21: ...ff the therapy device Note A therapy day starts and ends every day at 12 noon Data recorded from 0 00 midnight to 12 00 noon are always assigned to the previous calendar day 5 3 Auto On Off If Auto On...

Page 22: ...of therapy and correcting it Activate deactivate mask test Status therapy device is switched off 1 Call up the Patient menu and scroll to the setting for Mask test 2 Press the Softstart key or the Hu...

Page 23: ...ssed until the current setting appears 2 To change the time press the Softstart key or the Humidifier key several times until the desired time is displayed 3 To save the setting and exit the menu pres...

Page 24: ...atively with the therapy device switched on keep the Humidifier key depressed until the current setting appears 2 To change heating stage press the Softstart key or the Humidifier key several times un...

Page 25: ...idifier also switches back on at the heat output of the previous therapy 5 7 Operate therapy device Put on mask 1 Put on the mask as described in the section entitled 4 4 Breathing mask on page 16 2 C...

Page 26: ...rrent leakage displayed during therapy No leakage low leakage mask is perfectly positioned ensuring effective therapy Moderate leakage mask is not perfectly positioned quality of therapy may be restri...

Page 27: ...ly in the display 0 then appears in the display 3 Take the tube connection and the exhalation system off the breathing mask 4 Clean the following parts see 6 Hygiene treatment on page 29 breathing mas...

Page 28: ...ice In the carrying bag you should pack the therapy device the power cord the breathing tube the breathing mask incl exhalation system if appropriate the respiratory air humidifier replacement filters...

Page 29: ...hing mask and exhalation system See associated instructions for use Clean respiratory air humidifier See associated instructions for use Clean bacteria filter Replace particulate filter of bacteria fi...

Page 30: ...and outside with clean hot water 5 Shake out the breathing tube thoroughly 6 Hang up the breathing tube and allow it to drip dry 7 Dry the breathing tube with the aid of the therapy device Dry breath...

Page 31: ...start the drying process again 6 Remove the drying adapter if necessary 6 3 Clean housing Warning Risk of electric shock Before cleaning it is essential to disconnect the power cord from the power sup...

Page 32: ...uctions for use with regard to clean ing the exhalation system the breathing mask the headgear head strap the bacteria filter and the SOMNOclick 300 respiratory air humidifier 6 6 Disinfect therapy de...

Page 33: ...g proceed exactly as for cleaning Caution The breathing tube may not be heated to over 70 C Sterilization is not permitted Accessories For disinfecting sterilizing the exhalation system the breathing...

Page 34: ...nu set mask test pressure to 12 hPa see 5 4 Mask test on page 22 3 Switch on the therapy device The fan pumps air through the breathing tube to the breathing mask Current pressure is shown in the disp...

Page 35: ...age 21 Softstart cannot be switched on Sofstart function locked Clarify with your doctor whether Softstart can be enabled for your therapy Therapy device does not switch off about 5 seconds after mask...

Page 36: ...RTC now longer being sup plied with power Therapy can continue without restriction Inform your dealer Therapy device not working properly HF sources Increase the distance between devices emitting rad...

Page 37: ...off the click adapter 4 Push the new pressure measurement tube onto the click adapter 5 Hold up the creased tube and insert the free end of the new pressure measurement tube 6 Push the sleeve of the c...

Page 38: ...pter WM 24203 Carrying bag WM 23717 Power cord WM 24133 Coarse dust filter WM 24097 Fine filter packed WM 23596 Instructions for use WM 66941 Parts Order number SOMNOvent auto S WM 29205 SOMNOclick 30...

Page 39: ...ag WM 23717 Power cord WM 24133 Coarse dust filter WM 24097 Fine filter packed WM 23596 Instructions for use WM 66941 Parts Order number SOMNOvent auto ST WM 29405 SOMNOclick 300 WM 24375 Instructions...

Page 40: ...ing bag WM 23717 Power cord WM 24133 Coarse dust filter WM 24097 Fine filter packed WM 23596 Instructions for use WM 66941 Parts Order number SOMNOvent auto S 230 V WM 29210 SOMNOclick 300 WM 24375 In...

Page 41: ...ther breathing mask systems Parts Order number SOMNOvent auto ST 230 V basic device WM 29415 Tube system WM 24130 Drying adapter WM 24203 Carrying bag WM 23717 Power cord WM 24133 Coarse dust filter W...

Page 42: ...condensation Air pressure range 600 1100 hPa allows operation at up to 4000 m altitude Automatic altitude adaptation Connection diameter of breathing tube mask side in mm 19 5 fits standard size 22 mm...

Page 43: ...4 Parts 2 to 6 Part 11 Mean sound pressure level operation to EN ISO 17510 1 2002 at a distance of 1 m from the therapy de vice in the patient position approx 31 0 dB A at 18 hPa approx 29 0 dB A at 1...

Page 44: ...5 l min 180 l min 190 l min 15 l min Heating of respiratory air 2 5 C as per HMV Heilmit telverordnung German regu lations governing pharmaceutical products Depends on heating level Short term pressur...

Page 45: ...1 mbar 1 cm H2O 12 2 Pressure volume curve Long term pressure consistency to EN 17510 1 2002 p 0 2 hPa p 0 3 hPa Fine filter degree of separation up to 1 m up to 0 3 m 99 5 85 Fine filter service lif...

Page 46: ...n transmission frequency in m 150 kHz 80 MHz 80 MHz 800 MHz 800 MHz 2 5 GHz 0 01 0 1 0 04 0 07 0 1 0 37 0 11 0 22 1 1 2 0 35 0 70 10 3 7 1 11 2 21 100 11 7 3 50 7 00 O2 pressure source max 1 5 bar Flo...

Page 47: ...aim under warranty contact your specialist dealer 14 Declaration of Conformity L wenstein Medical Technology GmbH Co KG Kronsaalsweg 40 22525 Hamburg Germany the manufacturer of the devices described...