4 Function check
4 Function check
Carry out a function check at regular intervals:
• After installation
• After each update
• After changes to the complete system
4.1 Checking prismaTS
1. Start prismaTS.
2. Connect up a therapy device and all components (see
prismaTSlab instructions for use).
Requirement
: The therapy device is shown in the list of
detected devices.
3. Add a new patient.
Requirement
: New patient is saved.
4. Double-click on a therapy device in the list of detected devices.
5. Assign the newly added patient to the device.
Requirement
: The device is displayed for the patient.
6. Import the therapy data from the device.
If there are no therapy data available on the device, a therapy
must be performed first.
Requirement
: Therapy data are imported and displayed.
If one or more requirements are not satisfied or prismaTS is not
functioning correctly, please check that the therapy device and PC
are connected properly
. Restart the device and
prismaTS.