prisma CHECK
EN
39
5 Hygienic preparation
5.1 General information
• Wear suitable protective equipment for disinfection work.
• Please refer to the instructions for use supplied with the
disinfectant used.
5.2 Intervals
5.3 Hygienic preparation of the module
Requirement
The module is disconnected from the therapy device
Removing the module from the therapy device", page
1. Carry out hygienic preparation of the module as specified in
the following table:
2. Allow the module housing and in particular the electrical
interfaces to dry well.
Result
The module is hygienically prepared.
Interval
Action
Weekly
Clean the modules
As necessary, e.g., following infectious
diseases or severe soiling
Disinfect the modules
Damage to the device caused by ingress of liquids!
Ingress of liquids may damage the therapy device, components,
and accessories.
⇒
Do not immerse the therapy device, components, or
accessories in liquids.
Part
Cleaning
Disinfection
Sterilization
Housing
Wipe with a damp cloth
using water or mild soap
Wipe disinfection
(Recommendation: terralin
®
protect or perform advanced
Alcohol EP)
Not permitted
Connection cable