4 Preparation and operation
LM
T 6
85
41
a 0
6/2
02
1
EN
|
11
4 Preparation and operation
4.1 Setting up and connecting
device
1. If required: Tilt the device to a horizontal or vertical
position.
The display adapts to the orientation automatically.
2. Connect the power cable to the power supply unit and
the socket.
3. Connect the power supply unit cable to the device.
4.2 Connecting circuit
Risk of injury from inadequate therapy if air inlet and
air outlet are blocked!
A blocked air inlet and/or air outlet can cause the device to
overheat, impair therapy and damage the device.
Keep the air inlet clear.
Keep the filter compartment clear (
symbol).
Keep the outlet for the patient’s exhaled air fr
ee (
symbol).
Keep the intake opening for the cooling system free
(
symbol).
Material damage from overheating!
Excessive temperatures may lead to the device overheating
and damage the device.
Do not cover device and power supply unit with textiles
(e.g. bedclothes).
Do not operate device in the vicinity of a radiator.
Do not expose device to direct sunlight.
Only operate the device in the associated mobility bag
for mobile use.
Alternatively, you can connect a direct voltage
electricity supply (12 VDC or 24 VDC) as per
ISO 80601-2-72.
Risk of asphyxia if invasive or non-invasive patient/
ventilator interfaces without an exhalation system are
used!
If invasive or non-invasive patient/ventilator interfaces
without an integrated exhalation system are used, CO
2
concentration may rise to critical values and put the patient
at risk.
Use invasive or non-invasive patient/ventilator interfaces
with an external exhalation system if there is no
integrated exhalation system.
Follow the instructions for use for the exhalation
system.
Risk of injury from incorrectly routed circuit and
cables!
Incorrectly routed circuit or cables may injure the patient.
Do not route circuit and cables along the neck.
Do not crush circuit and cables.