A 5 | Ventilation modes and parameters
126
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
Reaching the pre-set SwitchFlow criterion
automatic release
pre-set
SwitchFlow criterion
automatic
release
PEEP
Flow
Paw
0
The SwitchFlow algorithm permanently considers the inspiratory and expirato
-
ry flow and checks their plausibility. Using this trigger function, which can be
switched off, the mandatory breaths can be synchronized.
A 5.5.23
PCV
Synchronised, time- and pressure-controlled ventilation mode with two adjust-
able pressure levels, allowing unimpeded spontaneous breathing on the lower
pressure level. Switching between both pressure levels guarantees a mandatory
minute volume and facilitates spontaneous breathing at PEEP level on the lower
pressure level, which increases the respiratory volume. Additional triggers start
additional mandatory breaths.