Ventilation modes and parameters | A 5
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
109
breathing.
The ventilator continuously monitors the patient's respiratory activity. If the min-
ute volume drops below the preselected value, a mandatory breath is initiated. If
spontaneous breathing fails completely, the patient is ventilated with the selected
mandatory respiratory rate.
A 5.5.9
Flexible BiLevel
Synchronised, pressure-controlled ventilation which allows the patient to breathe
spontaneously during the entire respiratory cycle. If the measured minute volume
is above the setting (Rate x Vt), the patient is able to breathe spontaneously. BiLev
-
el breaths are triggered with a dynamic adjustment of the mandatory ventilation
pressure if the minute volume drops below the setting.
With flexible BiLevel the advantages of pressure-controlled ventilation are com
-
bined with the possibility of unrestricted spontaneous breathing at both pressure
levels with a constant volume.
Depending on the current lung situation, the ventilator adapts the upper pressure
level or the necessary pressure support of spontaneous breathing so that the set
tidal volume is applied with the lowest possible inspiratory pressure.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...