A 5 | Ventilation modes and parameters
108
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
A 5.5.8
Optional BiLevel
Synchronised, pressure-controlled ventilation which allows the patient to breathe
spontaneously during the entire respiratory cycle. If the measured minute volume
is above the setting (Rate x Vt), the patient is able to breathe spontaneously. If the
minute volume drops below the setting, BiLevel breaths are triggered.
This mixed form of ventilation combines a synchronised, pressure-controlled, vol
-
ume-targeted ventilation mode to ensure a specified minute volume with the
possibility of achieving deeper breaths using pressure support.
Depending on the current lung situation, the ventilator adapts the mandatory
pressure level so that the set tidal volume is applied with the lowest possible
inspiratory pressure.
The Pmax and Pmin parameters determine a range in which the ventilator is
allowed to automatically readjust the necessary pressure to achieve the desired
tidal volume.
In contrast to conventional SIMV ventilation, the user defines a minimum minute
volume by setting the respiratory rate and tidal volume. No mandatory breaths
are initiated if the minute volume can be achieved by the patient's spontaneous
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...