Ventilation modes and parameters | A 5
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
103
A 5.5.3
Dual BiLevel ST
Dual BiLevel Spontaneous Timed.
Combination of BiLevel ST mode and dual BiLevel, allowing spontaneous breaths
during the entire respiratory cycle. The dynamic adaption of the ventilation pres-
sure refers to the mandatory and spontaneous portions of the respiratory cycle.
Breath cycles are defined by setting a fixed respiratory rate (60/f). If spontane
-
ous breathing results in the machine's trigger threshold being reached, pressure
support with volume guarantee (= dynamic PSV) will be provided for the spon
-
taneous breath. If the patient fails to breathe spontaneously during the cycle,
the machine will initiate a conventional BiLevel breath with volume guarantee
(= dynamic BiLevel) at the end of the calculated cycle time. In patients with no
spontaneous breathing activity, ventilatory support is delivered in the form of the
set mandatory minute volume (f x Vt). In patients who stop breathing for short
periods of time, the mandatory ventilation components provide an intelligent
back-up system for apnoea ventilation.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...