Weaninganalyzer
®
| A 4
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
87
A 4.5
P0.1
P0.1 occlusion pressure identifies the negative value which is established in the
first 100 milliseconds (0.1 s) of an inspiration phase (with closed valves) and rep
-
resents a measure for the neuromuscular respiratory drive of the spontaneously
breathing patient.
The values in a patient with healthy lungs who is breathing calmly are up to -3 and
–4 mbar. Values below -6 mbar indicate imminent respiratory fatigue.
The manoeuvre can be started in all ventilation modes. The only condition is the
patient being able to reach the trigger value.
The manoeuvre can be activated as a manual single measurement or for regular
monitoring at defined intervals (0.5 up to 120 minutes maximum). After the ma
-
noeuvre, the respective measured value (if configured) is displayed in the meas
-
urement value table.
Depending on the configuration, P0.1 can also be displayed and assessed in the
“Fastwean” display.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...