22
PCM_V2.1 12.2018
DANGER
DISCONNECT POWER BEFORE SERVICING! Only a qualified installation technician
should connect AC power to the PCM
1 .
Remove the 4 screws and front cover from the PCM .
2 .
Slide the Anybus module into position J100 on the low voltage circuit
board .
3 .
If necessary, field terminate the appropriate connector (DB9, RJ-45,
etc) and insert the connector into the Anybus module .
4 .
Replace the front cover and secure with 4 screws .
PCM Digital Interface
Description
The PCM digital interface uses Anybus
®
network modules to connect the PCM to a digital
network such as Profibus
®
DP or Modbus TCP . A block of registers is available for the
network master to scan, including measurements, alarms, errors and status information .
Installation
A single Anybus module is installed into the PCM as follows .
Figure 10:
Anybus connector
GB
Digital Interface