10
PCM_V1 07.2017
3 .
Remove the appropriate hole plugs from the bonding plate . The two
near the back are intended for high voltage use .
4 .
Connect AC power to the connector labeled J1 (see Figure 5 for
power connector location) . AC power and protective earth wires
must be 12 to 18 AWG . A local disconnect switch or circuit breaker
which meets local electrical code is also required on the AC power
input, and must be located near the PCM .
5 .
Connect Relay outputs to connectors labeled J2, J3, and J4 . Note that
Normally Open (NO) and Normally Closed (NC) contacts are provided/
See Figure 6 for wiring information .
6 .
Reinstall the low voltage board, connectors at position J3 and J9, and
the red thumb screw
GB
Installation
High Voltage Installation
DANGER
ELECTRIC SHOCK HAZARD . The barrier must stay in place except when a qualified
installation technician is connecting power or relay outputs
CAUTION
ENSURE PROPER GROUNDING . Connect metal bonding plate to protective earth
for proper operation
WARNING
Potential fire hazard . The relay contacts are not fused and rated at 5A max . External
devices should have a current limiting device to keep the current below 5A .