Lovibond SD 50 Manual Download Page 2

11. Set Temperature   
   System

Power ON

Press and hold (3 sec.)

Press 3 x

Temperature system will flash

10. Set Date and Time

Power ON

Select Date/Time type first (see 9.)

Press and hold (3 sec.)

Press 2 x

Time will flash
Confirm

Select minutes

To increase

To decrease

Save and change to hours

To increase

To decrease

Save and change to Year

To increase

To decrease

Save and change to Day/Month

To increase

To decrease

Save and exit

Replacement of batteries

pH Buffer

Article No.

72 12 50 pH buffer-set, each 90 ml   (25°C) 

  

4.00/7.00/10.00

72 12 52 pH buffer 4.00 (25°C)  

1 litre

72 12 54 pH buffer 7.00 (25°C)  

1 litre

72 12 56 pH buffer 10.00 (25°C)  

1 litre

51 56 00 buffer tablets pH 10  

100 pc.

51 56 01 buffer tablets pH 10  

250 pc.

51 56 10 buffer tablets pH 7  

100 pc.

51 56 11 buffer tablets pH 7  

250 pc.

51 56 20 buffer tablets pH 4  

100 pc.

51 56 21 buffer tablets pH 4  

250 pc.

Warranty

The waterproof SD 50 is warranted to be free 

from manufacturing defects for 2 years and 

electrode (probe) module for 6 months.

Error Code and Actions

Description 

Range Action

Electrode not installed  

Thermistor:  

Install or replace 

  

or faulty 

 0.5~100 k

Ω

 electrode

pH out of range 

0~14 pH 

Use new solution   

   

 

 

or replace electrode 

Electrode is not touching the   

 

 

Immerse electrode  

sample solution 

 

 

into solution

Temperature out of   

0~60 °C 

Bring solution to the  

range 

 

 

temperature within range 

Offset out of range  

-60~60 mV 

Use new buffer or     

    

 

 

replace  electrode

Slope out of range 

85%~115%,  

Use new buffer or     

50~68 (mV/pH) 

 

 

replace electrode

mV out of range  

-1800~1800 mV 

Use new solution  

SD 50 pH Specifications/

Features

1  Operating range: 0-60 °C, 0-14 pH,  

 

  -1800 mV to +1800 mV

2  pH resolution 0.01 pH. accuracy  

  ± 0,05 pH

3  Selectable buffer system (pH 7.00 or   

  pH 6.86): 1-, 2- or 3-point-calibration    

  with auto-recognition

4  Automatic Temperature compensation

5  mV resolution 0.1 mV within 

  ±1000 mV and 1 mV over 

  ±1000 mV, accuracy ±20 mV

6  Temperature resolution and accuracy  

  0.1 °C, selectable °C or °F

7  Time and date display with 25 sets of    

  data storage (non-volatile)

8  22x22 mm LCD display, with backlight

9  2 x CR 2032 batteries

10 Battery life > 25 hr (continuous use,  

 

  backlight OFF), low battery indicator on   

  LCD display

11 Auto-power off to save battery-life  

  (8 minutes non-use)

Instruction manuals available in further languages, visit www.lovibond.com • Printed in Germany 11/13

Storage of electrode 

The sponge in the protective cap should be 

moistened with tap water.
Store electrodes dry when instrument is not 

being used for longer periods. 
Moisten electrode in tap water for some hours 

before initial use and after dormant periods. 

1

2

3

4

5

6

Replacement of electrode

new electrode

Confirm

Select Temperature style (°F/°C)

Save and exit

Accessories

19 50 017   Battery CR 2032 (2 Batteries    

  

required)

19 48 20   pH replacement electrode

38 48 01   Beaker, 100 ml

+

+

+ +

-

-

-

-