3.6 Troubleshooting
3.6.1 Operating messages in the display / error display
Display
Possible Causes
Elimination
Overrange
reading is exceeding the range
water sample is too cloudy
to much light on the photo cell
if possible dilute sample or
use other measuring range filtrate
water sample
seal on the cap?
Repeat measurement with seal on the
cap of the vial
Underrange
result is under the detection limit indicate result with lower x mg/l
x = low end of measuring range;
if necessary use other analytical
method
Storage-
system error
use Mode 34
mains power fails or is not
connected
insert or change Lithium battery
Delete Data with Mode 34.
Battery
warning
warning signal every 3 minutes
warning signal every 12 seconds
capacity of the battery is too low;
change the batteries
warning signal, the instrument
switches itself off
change the batteries
Jus
Overrange
E4
The user calibration is out of the
accepted range
Please check the standard, reaction
time and other possible faults.
Repeat the user calibration.
Jus
Underrange
E4
Overrange
E1
The concentration of the standard
is too high/too low, so that
during user calibration the limit
of the range was exceeded
Perform the test with a standard of
higher/lower concentration
Underrange
E1
E40 user
calibration
not possible
If the display shows Overrange/
Underrange for a test result a
user calibration is not possible
Perform the test with a standard of
higher/lower concentration
Summary of Contents for PoolDirect
Page 2: ......
Page 4: ......
Page 12: ...12 PoolDirect 9in1_4i 11 2015...
Page 13: ...13 PoolDirect 9in1_4i 11 2015 Part 1 Methods...
Page 35: ...35 PoolDirect 9in1_4i 11 2015 Part 2 Operator manual...
Page 38: ...38 PoolDirect 9in1_4i 11 2015...
Page 70: ...70 PoolDirect 9in1_4i 11 2015...
Page 71: ...71 PoolDirect 9in1_4i 11 2015 Part 3 Enclosure...