One Time Zero (OTZ)
OneTimeZero is available for all methods where zero is performed in a 24 mm Ø round vial
with sample water (see chapter 1.1 Table of Methods).
OneTimeZero can be used for different tests providing the tests are performed with the same
sample water and under the same test conditions. When changing the method, it is not
necessary to perform a new Zero. The test can be carried out straight away.
When the instrument is first being used for an OTZ compatible method and OneTimeZero is
activated, the instrument will request a new Zero with “prepare OT-Zero”. Perform Zero as
described in the method. This Zero will be stored and used for all methods with OTZ function
until the instrument is switched off.
If necessary, a new Zero can be performed by pressing [Zero] key at any time.
Switching the “OTZ-Function“ on and off:
Press [MODE] [5][5] keys in sequence.
Confirm with [ ] key.
The display shows:
• Press [0] key to switch the OTZ off.
• Press [1] key to switch the OTZ on.
The display shows:
Confirm with [ ] key.
The instrument goes back to mode menu automatically.
Note:
The specified accuracy is valid for all test results when Zero is performed for each test
(OneTimeZero function is switched off).
2.4.7 blank because of technical requirements
Mode
5
5
<OneTimeZero>
ON : 1 OFF : 0
1
0
switched off
or
switched on
Summary of Contents for PoolDirect
Page 2: ......
Page 4: ......
Page 12: ...12 PoolDirect 9in1_4i 11 2015...
Page 13: ...13 PoolDirect 9in1_4i 11 2015 Part 1 Methods...
Page 35: ...35 PoolDirect 9in1_4i 11 2015 Part 2 Operator manual...
Page 38: ...38 PoolDirect 9in1_4i 11 2015...
Page 70: ...70 PoolDirect 9in1_4i 11 2015...
Page 71: ...71 PoolDirect 9in1_4i 11 2015 Part 3 Enclosure...