61
PoolDirect 9in1_4i 11/2015
Store user calibration
Perform the required method as described in the manual
using a standard of know concentration instead of the
water sample.
If the test result is displayed press [MODE] [4] [5] keys in
sequence and confirm with [ ] key.
The display shows:
Pressing the arrow key [ ] once increases the displayed
result.
Pressing the arrow key [ ] once decreases the displayed
result.
Press keys as long as the displayed result corresponds to the
value of the standard.
Confirm with [ ] key to store the new calibration factor.
Cancel user calibration by pressing [ESC] key.
The display shows:
Now the method name is displayed inverse and the test result
is calculated with the new calibration factor.
Mode
4
5
Chlorine T
0.01-6 mg/l Cl2
0.90 mg/l free Cl2
<user calibration>
Chlorine T
0.01-6 mg/l Cl2
0.90 mg/l free Cl2
up: , down:
save:
Jus Factor
saved
Chlorine T
0.01-6 mg/l Cl2
1.00 mg/l free Cl2