Chlorine L / M101
EN Method Reference Book 1.0
23
EN
Zero
Press the
ZERO
button.
Remove the vial from the
sample chamber.
Empty vial.
6
2
Hold cuvettes vertically
and add equal drops by
pressing slowly.
Add
6 drops DPD 1 Buffer
Solution
.
Add
2 drops DPD
1 Reagent Solution
.
3
Add
3 drops DPD
3 Solution
.
Fill up vial with
sample
to
the
10 ml mark
.
Close vial(s).
Sample
Test
Invert several times to mix
the contents.
Place
sample vial
in the
sample chamber. • Pay
attention to the positioning.
Press the
TEST
(XD:
START
)button.