81
MaxiDirect_2c 03/2010
1.1 Methods
Notes:
1. Application: Testing of residual corrosion inhibitors (Oxygen scavengers) in boiler feed
water or condensate.
2. Before using clean the vials with Hydrochloric acid (approx. 20%). Rinse thoroughly with
deionised water.
3. Keep the sample dark during colour development time. UV-light (sunlight) causes high
measurement results.
4. Ideal temperature for full colour development is 20°C ± 2°C.
5. Interferences:
• Iron (II) interferes at all concentrations:
Repeat the test procedure but without adding the DEHA solution. If the
displayed result is above 20 μg/l subtract this value from the DEHA test result.
• Substances which reduce Iron (III) interfere. Substances which complex iron strongly
may interfere also.
• Substances which may interfere when present in concentrations at:
Borate (as Na
2
B
4
O
7
)
500 mg/l
Cobalt
0.025 mg/l
Copper
8.0 mg/l
Hardness (as CaCO
3
)
1000 mg/l
Lignosulfonates
0.05 mg/l
Manganese
0.8 mg/l
Molybdenum
80 mg/l
Nickel
0.8 mg/l
Phosphate
10 mg/l
Phosphonates
10 mg/l
Sulfate
1000 mg/l
Zinc
50 mg/l
6. There is an option to change the unit from mg/l to μg/l.
mg/l
μg/l
Summary of Contents for Maxi Direct
Page 1: ...Photometer System axi GB Instruction manual ...
Page 2: ......
Page 9: ...7 MaxiDirect_2c 03 2010 Part 1 Methods ...
Page 115: ...113 MaxiDirect_2c 03 2010 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 143: ...141 MaxiDirect_2c 03 2010 ...
Page 193: ...191 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 206: ...204 MaxiDirect_2c 03 2010 ...
Page 207: ...205 MaxiDirect_2c 03 2010 Part 2 Instrument Manual ...
Page 210: ...208 MaxiDirect_2c 03 2010 ...
Page 258: ...256 MaxiDirect_2c 03 2010 ...
Page 259: ...257 MaxiDirect_2c 03 2010 Part 3 Enclosure ...
Page 267: ...265 MaxiDirect_2c 03 2010 ...