76
MaxiDirect_2c 03/2010
1.1 Methods
Cyanide
with Reagent Test
0.01 – 0.5 mg/l CN
1. Fill a clean vial (24 mm Ø) with
2 ml of the water
sample and 8 ml of deionised water
, close tightly
with the cap.
2. Place the vial in the sample chamber making sure that
the marks are aligned.
3. Press
ZERO
key.
4. Remove the vial from the sample chamber.
5. Add
two level spoons No. 4 (grey) of Cyanide-11
into the prepared water sample, replace the cap tightly
and invert the vial several times to mix the contents.
6.
Add two level spoons No. 4 (grey) of Cyanide-12
,
replace the cap tightly and invert the vial several times
to mix the contents.
7. Fill the vial with drops of the same size by holding the
bottle vertically and squeeze slowly:
3 drops of Cyanide-13
8. Close the vial tightly with the cap and invert several
times to mix the contents.
9. Place the vial in the sample chamber making sure that
the marks are aligned.
10. Press
TEST
key.
Wait for a reaction
period of 10 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l Cyanide.
Countdown
10:00
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
5
7
1
Summary of Contents for Maxi Direct
Page 1: ...Photometer System axi GB Instruction manual ...
Page 2: ......
Page 9: ...7 MaxiDirect_2c 03 2010 Part 1 Methods ...
Page 115: ...113 MaxiDirect_2c 03 2010 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 143: ...141 MaxiDirect_2c 03 2010 ...
Page 193: ...191 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 206: ...204 MaxiDirect_2c 03 2010 ...
Page 207: ...205 MaxiDirect_2c 03 2010 Part 2 Instrument Manual ...
Page 210: ...208 MaxiDirect_2c 03 2010 ...
Page 258: ...256 MaxiDirect_2c 03 2010 ...
Page 259: ...257 MaxiDirect_2c 03 2010 Part 3 Enclosure ...
Page 267: ...265 MaxiDirect_2c 03 2010 ...