28
MaxiDirect_2c 03/2010
1.1 Methods
Ammonia HR
with Vario Tube Test
1 – 50 mg/l N
Insert the adapter for 16 mm Ø vials.
1. Open one white capped reaction vial and add
0.1 ml
deionised water
(this is the blank).
2. Open another white capped reaction vial and add
0.1 ml of the water sample
(this is the sample).
3. Add the contents of
one Vario Ammonia Salicylate
F5 Powder Pack
straight from the foil into each vial.
4. Add the contents of
one Vario Ammonia Cyanurate
F5 Powder Pack
straight from the foil into each vial.
5. Close the vials tightly with the caps and swirl several
times to dissolve the powder.
6. Press
[ ]
key.
Wait for a
reaction period of 20 minutes
.
After the reaction period is finished proceed as follows:
7. Place the vial (the blank) in the sample chamber making
sure that the marks are
l
aligned.
8. Press
ZERO
key.
9. Remove the vial from the sample chamber.
10. Place the vial (the sample) in the sample chamber
making sure that the marks are
l
aligned.
11. Press
TEST
key.
The result is shown in the display in mg/l Ammonia as N.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown 1
20:00
start:
1
2
3
4
Ø 16 mm
6
6
Summary of Contents for Maxi Direct
Page 1: ...Photometer System axi GB Instruction manual ...
Page 2: ......
Page 9: ...7 MaxiDirect_2c 03 2010 Part 1 Methods ...
Page 115: ...113 MaxiDirect_2c 03 2010 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 143: ...141 MaxiDirect_2c 03 2010 ...
Page 193: ...191 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 206: ...204 MaxiDirect_2c 03 2010 ...
Page 207: ...205 MaxiDirect_2c 03 2010 Part 2 Instrument Manual ...
Page 210: ...208 MaxiDirect_2c 03 2010 ...
Page 258: ...256 MaxiDirect_2c 03 2010 ...
Page 259: ...257 MaxiDirect_2c 03 2010 Part 3 Enclosure ...
Page 267: ...265 MaxiDirect_2c 03 2010 ...