213
MaxiDirect_2c 03/2010
2.3.5 Performing Tests
When zero calibration is complete, remove the vial from the sample chamber and perform
the tests as described under “Method“.
When the results have been displayed:
- with some methods you can change between different chemical species
- you can store and/or print out the results
- perform further analysis with the same zero
- select a new method
2.3.6 Ensuring reaction periods (countdown)
To ensure compliance with reaction periods a time delay is incorporated: the countdown.
There are two kinds of countdowns:
• Press [
] key.
Prepare water sample, start countdown with [ ] key and
proceed as described in the mode description.
The vial must not be placed in the sample chamber.
• Press [TEST] key.
Prepare the water sample as described in the method
description and place the vial in the sample chamber.
The display shows the countdown by pressing the [TEST]
key and the countdown is started automatically. After
the reaction period is finished the measurement starts
automatically.
Notes:
1. It is possible to finish the working countdown by pressing the [ ] key. Reading starts
immediately. In this case the operator is responsible for ensuring the necessary reaction
period.
Non-compliance with reaction periods leads to incorrect test results.
2. The time remaining is displayed continuously.
The beeper indicates the last 10 seconds.
Countdown
1:59
Countdown
2:00
start:
Test
Summary of Contents for Maxi Direct
Page 1: ...Photometer System axi GB Instruction manual ...
Page 2: ......
Page 9: ...7 MaxiDirect_2c 03 2010 Part 1 Methods ...
Page 115: ...113 MaxiDirect_2c 03 2010 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 143: ...141 MaxiDirect_2c 03 2010 ...
Page 193: ...191 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 206: ...204 MaxiDirect_2c 03 2010 ...
Page 207: ...205 MaxiDirect_2c 03 2010 Part 2 Instrument Manual ...
Page 210: ...208 MaxiDirect_2c 03 2010 ...
Page 258: ...256 MaxiDirect_2c 03 2010 ...
Page 259: ...257 MaxiDirect_2c 03 2010 Part 3 Enclosure ...
Page 267: ...265 MaxiDirect_2c 03 2010 ...