121
MaxiDirect_2c 03/2010
1.1 Methods
Notes:
1. Filter turbid water samples using filter paper and funnel before analysis.
2. Highly buffered water samples or extreme pH values should be adjusted to a pH of
nearly 7 with 1 mol/l Nitric acid or 1 mol/l Sodium hydroxide.
3. Concentrations above 10 mg/l Cu causes too high test values if the reaction time of
5 minutes is increased. So it is very important to perform the test procedure
as described.
4. Substances which may interfere when present in concentrations at:
Aluminium
50 mg/l
Chromium
1000 mg/l
Iron
50 mg/l
Nickel
50 mg/l
Nitrite
all levels
5. MoO
4
Mo
Na
2
MoO
4