1.1 Methods
Hydrazine
with Vacu-vials
®
K-5003
(see Notes)
0.01 – 0.7 mg/l N
2
H
4
/ 10 – 700 μg/l N
2
H
4
Insert the adapter for 13 mm Ø vials.
1. Place the blank in the sample chamber. The blank is part
of the test kit.
2. Press
ZERO
key.
3. Remove the blank from the sample chamber.
4. Fill the sample container to the 25 ml mark with the
water sample.
5. Place one Vacu-vial
®
in the sample container. Snap the
tip by pressing the vial against the side of the sample
container. The Vacu-vial
®
breaks at the neck and the vial
fills automatically. A small volume of inert gas remains
in the Vacu-vial
®
.
6. Mix the contents of the Vacu-vial
®
by inverting it several
times, allowing the bubble to move from one end to
the other. Dry the outside of the vial.
7. Place the Vacu-vial
®
in the sample chamber.
8. Press
TEST
key.
Wait for a
reaction period of 10 minutes
.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display as Hydrazine.
Countdown
10:00
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
0
7
2
Summary of Contents for Maxi Direct
Page 1: ...Photometer System axi GB Instruction manual ...
Page 2: ......
Page 9: ...7 MaxiDirect_2c 03 2010 Part 1 Methods ...
Page 115: ...113 MaxiDirect_2c 03 2010 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 143: ...141 MaxiDirect_2c 03 2010 ...
Page 193: ...191 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 206: ...204 MaxiDirect_2c 03 2010 ...
Page 207: ...205 MaxiDirect_2c 03 2010 Part 2 Instrument Manual ...
Page 210: ...208 MaxiDirect_2c 03 2010 ...
Page 258: ...256 MaxiDirect_2c 03 2010 ...
Page 259: ...257 MaxiDirect_2c 03 2010 Part 3 Enclosure ...
Page 267: ...265 MaxiDirect_2c 03 2010 ...