5.9 Error messages and notices
Message in the display
Meaning
RTC Error!
Please set the date and time
again.
The message appears when the device is switched on and the RTC
(real-time clock) has forgotten its time. Adjust the time in the follo-
wing menu and then check all pending measurements.
LOBAT!
This appears just before the device switches off on its own. Replace
the batteries.
USB-Host overcurrent!
Remove the USB device.
This message appears if a short-circuit has occurred at the USB host
port. Remove the USB device.
Incorrect head ID!
The measurement series was
not started!
This appears at the start of a new measurement series when a diffe-
rent head was found at the selected measurement position than for
the reading of measurement parameters.
Communication error!
The measurement series was
not started!
This appears when a new measurement series is started, if there is
no head available at the selected measurement position.
The battery level is low!
This appears when a new measurement series is started, if the
batteries are nearly depleted. Replace the batteries or connect the
mains adapter. The measurement series can still be started.
The current measurement
series will be overwritten!
The current measurement series will be overwritten!
Internal processing error!
The measurement series was
not started!
This appears when a new measurement series is started, if the
displayed measurement parameters cannot be correctly interpreted.
Change the measurement parameters back to the previous parame-
ters. Then restart measurement series.
Start error!
Please restart the measure-
ment series!
This appears when a new measurement series is started, if the mea-
surement could not be started. Check the contacts at the head and
the measurement position. Restart the measurement series.
The measurement series was
started.
Note. This appears if a measurement series was started successfully.