21
BD 600_GB_1 08/2015
5.6.9 Auto-Off
The device can be switched off independently after three, five, or ten minutes.
The background lighting is already switched off ahead of time. The times are always based on the last
actuation of a key/key. If auto-off is deactivated, the device does not switch off independently.
5.6.10 Remote control
If this item is switched on, the device reacts to the remote control. Otherwise, the remote control is
ignored. If ‚Mains operation‘ is set here, the remote control does not work in battery operation only.
5.6.11 Device ID
The device ID is set for the remote control here, with which the device can be selected (see also chap-
ter 5.8 Remote control).
5.6.12 Update
In this field, the current version number of the operating software can be seen.
You need the ‚bod.hex‘ file on a USB stick or an SD card for an update. In the process, the file must
be in the root directory of the data carrying medium - in other words, it may not be in a subfolder,
otherwise it will not be found by the device.
If both an SD card and a USB stick with such a file are inserted, the file on the SD card is used.
For security reasons, updating is only possible in mains operation, so depleted batteries cannot inter-
fere with the update.
Ensure that both conditions are met and start the update with the right arrow
►
key. A security query
follows. Confirm this with function key F2 (OK). Now the boot loader which performs the update
starts up. The device then restarts.
Updates can be found on our website www.lovibond.com under ‚Support‘.