V-Link
®
-200
User Manual
3.5
Configure Node
Node settings are stored to non-volatile memory and may be configured using SensorConnect. To
access the node configuration menu, under Devices select the node and then the Configure tile.
The configuration menus show the channels and configuration options available for the type of node
being used.
For this example the V-Link-200 tester board is on channel 1.
1. Select Hardware > Input Range for channel 1, /-2 mV from the drop down menu.
2. Under Hardware Offset, select Balance Target for channel 1, select Mid (50%) from the drop
down menu.
3. Select Auto- Balance. When auto- balance is complete, a blue information window will
indicate the balance result.
Figure 16 - Auto-Balance
4. Select Calibration.
5. Select Microstrain from the Unit drop down menu, and select the Shunt Cal button enabled
on the right.
23