P/N 9-500-0255
4-7
Service Manual
Chapter 4: Calibrations and Performance Checks
Compression System Calibration Procedures
3.0
Compression System Calibration Procedures
3.1
Compression Force Calibration
This procedure calibrates the compression force sensing circuits for accurate force display.
Be sure that the load cell selected is the type of load cell installed (refer to Table 4-2). Note
that the load cell type is factory set and should never be changed unless the load cell is
replaced.
1. Turn the unit OFF and remove the right side Gantry cover to access the Microprocessor
Board. Set the machine for the service mode (S2 switch 1 = ON).
2. Apply power. When the Run Mode appears, access Calibration Screen #3 by pressing
Reset
+
Left Arrow
.
3. Remove any installed accessory from the Image Receptor Support Device.
4. Place a force gauge (or scale) on the IRSD. Center the scale laterally and position it 1.0
cm from chest wall edge. Install the 18 x 24 cm screening compression paddle onto the
compression device.
5. Highlight the CF Cal data field using the
Scroll
keys. Press a
Change
key until the field
reads: 80 N (18 lb).
6. Place a towel on the gauge
or scale to protect the
compression accessory
from scratches or damage.
Use motorized compres-
sion to lower the compres-
sion device until the
paddle just touches the
force gauge.
7. Use the manual compres-
sion handwheels to apply
additional compression
until the gauge or scale
indicates 18 lb of force.
Press
Reset
+
Comp.
Release
together to send
the 18 lb calibration data
to the microprocessor.
Figure 4-1: Compression Force Calibration Setup
8. Press a
Change
key until the data field reads: 151 N (34 lb). Apply additional manual
compression until the gauge or scale indicates 151 N (34 lb) of force.
9. Press
Reset
+
Comp. Release
together to send the 151 N (34 lb) calibration data to the
microprocessor.
10. Raise the compression device off of the gauge or scale, then remove the towel and
gauge. Perform the Compression Thickness calibration as per Section 3.2.