System Specifications
Figure A-1: Unit Measurements
1.3
Operating Environment
Temperature Range
10° C to 40° C (50° F to 104° F), IEC 601-1,
Par. 10.2.1A
Relative Humidity Range
30% to 75% non-condensing, IEC 601-1, Par.
10.2.1B
ESD Susceptibility
Level of 3 kV for contact discharge to conduc-
tive accessible parts that are not grounded. A
level of 8 kV for air discharge to all accessible
parts. IEC 601-1-2
EMI Susceptibility
System is immune from levels of 1 v/m for the
frequency range of 26 MHz to 1 GHz. IEC
601-1-2
EMI Generation Limits
System complies with the requirements of IEC
601-1-2 for conducted and radiated emission.
Input Line
Surge, fast transient/burst, lightning, IEC 601-
1-2
Maximum Heat Dissipation
1700 Btu/Hour in stand-by
1.4
Storage Environment
Temperature Range
-25° C to +60° C (-13° F to +140° F)
Humidity
0 to 95% (non-condensing); not packaged for
outdoor storage