PyroTec™ PRO Hardw are User’s Manual, V1.2
83
RT-MN024-1.2 02/21
Use
: Decon / Contrad for surface cleaning only
Do not use
: Decon / Contrad, Bleach, Peraclean, SporGon as clean-
ing bath for carriers (damage to aluminum)
Racks
Water, alcohol, weak detergent, disinfectant
Gripper
Water, alcohol, weak detergent, disinfectant
Safety panels
Water, alcohol, disinfectant, suitable for acrylic glass
Disposable tip cones
Alcohol
Arm guide, arm guide
roller of arms
Do not use any agent
Z-rod
Do not use any agent
Note
: After use of weak detergents, base or bleach, thoroughly clean with water and wipe
dry to totally remove the cleaning agent and obtain normal operating conditions.
Cleaning
Tissue
Cleaning
Use a lint-free tissue together with the appropriate cleaning agent.
9.1.2 For Air LiHa Maintenance
Material
Needed for
Maintenance
To perform the maintenance on the Air LiHa, the following tools and consumables must be
available:
•
Special tools and consumables
o
Inline Filter kit Air LiHa (incl. filter removal tool)
o
DiTi cone wrench
o
Air LiHa tip adapter adjustment gauge (optional)
9.2 Maintenance Schedule
Note
: To ensure a good working condition of the PyroTec instrument, a yearly (depending
on configuration) maintenance carried out by an authorized field service engineer is rec-
ommended.
Maintenance
Record
Note
: In order to be able to track all maintenance performed on the PyroTec™ PRO over
the whole lifetime, the periodic maintenance must be recorded as follows:
•
Fill in the necessary data in the form “Daily/Weekly Maintenance Checklist”.
•
File the form in the “Maintenance and Service Logbook”.
Maintenance
Tables
The maintenance tables are divided according to the frequency the corresponding mainte-
nance task must be periodically performed. For example, there are tables for:
•
Daily maintenance
•
Weekly maintenance
•
Half-yearly maintenance