PyroTec™ PRO Hardw are User’s Manual, V1.2
74
RT-MN024-1.2 02/21
In each rack, use tubes of one size only. Tube height and diameter must be identical for all
tubes.
ATTENTION
Ensure all tubes are positioned correctly in the carrier touching the rack bottom; other-
wise liquid level and clot detection functions might not work properly.
Minimum inner diameter for primary sample tub es
Tip type
Tube diameter, inside
DiTi 1000
µl
8 mm
8.3.4.3
Preparation of Samples
ATTENTION
Visually inspect the samples before pipetting They must be free of:
•
Clots
•
Form
•
Droplets on the tube wall
Droplet on wall
•
The sample tubes must not contain any additional (non-conductive) inserts or have co-
vers.
Note
: For further information on sample preparation, please refer also to the recommen-
dations given by your manufacturer and by the WHO.
8.3.5 Runtime Controller
ATTENTION
The runtime controller of the application software controls The PyroTec™ PRO.
With the runtime controller, the following tasks are performed during installation and
maintenance where appropriate. Runtime controller should only be accessed by Lonza ap-
proved engineers: