PyroTec™ PRO Hardw are User’s Manual, V1.2
68
RT-MN024-1.2 02/21
In the event of power failure or an otherwise aborted run, all partially processed samples
should be discarded. Do not attempt to restart an interrupted program unless the com-
puter screen displays explicit instructions for resuming operation.
Chemical, Biological and Radioactive Hazards
WARNING
All samples and test kit components must be considered potentially hazardous agents.
•
A potential risk can arise from the liquids being handled by the instrument, such as in-
fectious biological samples, toxic or corrosive chemicals, or radioactive substances.
•
Strictly apply appropriate safety precautions according to local, state, and federal reg-
ulations.
•
Handling and disposing of waste must be in accordance with all local, state, and fed-
eral environmental, health, and safety laws and regulations.
•
Use appropriate protective clothing, safety goggles and gloves.
8.3.2 Enclosed Work Area
WARNING
Unexpected, fast movements of arms and tips.
Interfering with the arm and tip movements can lead to serious injuries or equipment
damaging.
Never operate the instrument while safety panels, covers or access doors are open or re-
moved.
The software will prompt the operator when the worktable setup requires new racks or
carriers. Any further interference in the work area is strictly prohibited.
The operator might need to open or remove the safety panels for instrument setup, clean-
ing, and maintenance purposes.
8.3.3 Switching the Instrument On
Before switching the instrument on, checking the following:
WARNING
If the pipetting head is initialized with the tip block or DiTis mounted to it, the remaining
liquid in the tips may be drawn into the head and thus contaminate the head (e.g. after a
power failure or crash).
In this case, manually remove the tip block or the DiTis before switching on.