Operating Instructions
To perform a ferrite tune, proceed as follows:
1.
Select
Menus > Service > Options
to display the
Service Options
menu options which are
listed and described below.
2.
Use the
Navigation Keys
to select the
Access Ferrite
option and press
the
OK / Enter Key
to select it
3.
Use the
Navigation Keys
to display the
YES
option and then press the
OK
/ Enter Key
to select it.
4.
Navigate to the
Menus > Change
menu, then use the
Navigation Keys
to
display the
Ferrite F
product and press the
OK / Enter Key
to select it.
5.
Perform a Full Learn whilst moving the end of the ferrite wand with the sample in and out of
the search head; taking care not to put your hand in or get any watches/rings near the search
head.
On systems fitted with a Product Registration Photo-eye, block the photo-eye to register that
packs are present.
6.
Once a successful Learn has been completed, navigate back to the
Menus > Service >
Options
menu and de-select the
Access Ferrite
option.
You have now completed the Ferrite Tune process.
Setting Up Product Signal Tracking
General
This feature allows the detector to automatically adjust the working threshold between two pre-set
limits to enable the sensitivity of the detector to be optimised in situations where the product signal
drifts up and down with the change in product effect.
For instance, for a frozen product the product effect is very small when the product has just come out
of the freezer allowing a low working threshold level to be used giving a high sensitivity to metal
contaminants. However, if the line stops for any reason then the product could thaw slightly, or if the
product were not fully frozen, a significant product signal increase would result. In this event the
working threshold will have to be raised to allow the product to be passed, without false rejection.
Raising the working threshold to allow softer product to pass without false rejection will reduce the
sensitivity to metal contamination, i.e. larger pieces of metal will now be detected. If the product
becomes hard frozen whilst the raised working threshold is still set the sensitivity to metal
contamination will remain reduced because the working threshold is higher than it now needs to be.
Tracking works by recording the product signal for each pack. From the values read the working
threshold continuously trends upwards or downwards between an upper and a lower limit.
The rate and amount by which the working threshold changes, together with limits between which it
operates, are controlled by user entered parameters.
It must be understood that this type of system may allow metal contamination that would otherwise be
detected when the product is fully frozen, to be accepted if product softening has forced the working
threshold to rise.
88
Summary of Contents for IQ3
Page 2: ......
Page 10: ......
Page 26: ......
Page 43: ...Installation 33 ...
Page 60: ...Installation THIS PAGE IS INTENTIONALLY LEFT BLANK 50 ...
Page 70: ...About the Metal Detector Range THIS PAGE IS INTENTIONALLY LEFT BLANK 60 ...
Page 74: ...Getting Started IQ3 Dual Line Display Menu Map 64 ...
Page 116: ...Operating Instructions THIS PAGE IS INTENTIONALLY LEFT BLANK 106 ...
Page 140: ...Maintenance THIS PAGE IS INTENTIONALLY LEFT BLANK 130 ...