background image

This quickstart guide covers preparation, 

use and maintenance for the following model: 

UTRID-16

1. SOFTWARE SETUP

2. LOGTAG

®

 CONFIGURATION

Software requirements vary depending on the type of file you want it to 
generate when plugged into a USB port.

Loggers can be ordered pre-configured, ready to start. Different 
profiles are available; if none of these suit or the logger needs different 
configuration settings to those already installed, you will need to 
download and install our freely available companion LogTag

®

 Analyzer 

(version 2.8 or higher) software from our website: www.logtagrecorders.
com/software. Simply browse to our software page, fill in your details and 
begin your download.

Once it has finished downloading, install LogTag

®

 Analyzer to your 

computer and start the program once it is installed. (For detailed 
software download and install instructions, please refer to the LogTag

®

 

Analyzer User Guide).

It is recommended that you connect each USB PDF Temperature logger 
via a permanently plugged in USB extension cable to protect your 
computer’s USB socket. Please refer to 
www.logtagrecorders.com/support if you are having issues.

With the Software running, the easiest way to configure your LogTag

®

 

for use is with the ‘

LogTag Wizard

’ which can be accessed by 

pressing the ‘

F2

’ key on your keyboard or you can browse via the top 

navigation menu: ‘

LogTag

’ 

>

 ‘

Wizard

Note: 

You can configure many devices at the same time, however, 

it is practical to limit the number of devices to about 10, using a 
powered USB hub. It is not possible to configure different models at 
the same time.

Advanced Configuration Options

Click 

Advanced Settings

 for additional configuration settings. 

These settings decide which files -if any- are generated at the end 
of the trip and also determine the appearance and contents of the 
generated files.

Alarm Configuration

You need to observe some basic rules when entering alarm 
conditions into the configuration screen in LogTag

®

 Analyzer:

• 

A primary alarm condition must be entered before the 
corresponding secondary alarm condition.

• 

Equally, primary and secondary alarm conditions must be entered 
before the corresponding tertiary alarm condition.

• 

You can enter a different number of upper and lower conditions, 
or only upper, or only lower conditions, or none at all.

• 

A secondary upper alarm cannot be entered with a lower 
threshold value than the primary upper alarm.

• 

A tertiary upper alarm cannot be entered with a lower threshold 
value than the secondary alarm. Threshold values can, however, 
be equal and combined with different time-to-activation values. 

 

For example, you can enter a primary upper alarm with 8°C 
threshold and 10 accumulative readings, plus an alarm with 8°C 
threshold and 5 consecutive readings. In this instance the alam 
will be triggered, if either 10 readings in total are above 8°C, or 5 
consecutive readings above 8°C have occurred.

Any alarm condition that is activated will be shown on the PDF, in the 
software and on the display.

“Standard Configuration”

UTRID-16 Quickstart Guide

Page 1 of 3

1.  Provide a ‘Description’ to identify the logger being configured.

2.  Select Push button or Time/Date start.

3.  Enter the recording interval you wish to use. The duration  

of the shipment and/or number of readings to record changes 
according to the recording interval entered. 

4.  (Optional) Enter a start delay before the logger begins recording.

5.  LogTag

®

 Analyzer allows up to 6 different alarm trigger conditions 

when configuring a UTRID-16 for recording, depending on the 
model. Each alarm trigger condition consists of a threshold 
temperature value, an activation type (which can be instant, 

1

2

3

4

consecutive or accumulative) and a delay time, if it is not an 
instant alarm. The UTRID-16 can have up to three upper alarms, 
and up to three lower alarms.

5

Reviews: