Maintenance
5.1 General care
Clean surfaces of the main unit and controller with a soft cloth dampened with distilled
water or 70% ethanol. Immediately wipe dry with a clean cloth.
Do not pour or spray liquids directly onto the instrument.
To avoid electrical shock or damage, do not wet electrical wires or connections.
If liquid is spilled on the instrument, turn off the power and wipe dry immediately.
Do not exchange components between instruments unless they have been provided or
authorized by Logos Biosystems.
Use only optical-grade cleaning materials to clean optical components.
Do not use abrasive or sharp-edged devices when cleaning the system.
5.2 Change filter cubes
Procedure
Go to Settings > Filter Cubes.
Click Change filter cubes.
Click Start.
Remove the vessel holder from the stage.
Remove the filter cube stage cover.