Maintenance
Battery Replacement
1. Remove the single screw on the battery door, located in the back of the
FaultChaser
towards the bottom of the unit
, with a #1 Philips head screwdriver.
2. Take off the battery door and remove old batteries.
3. Replace with four AA Alkaline batteries. Slide batteries into the battery cartridge
according to the diagram imprinted on the bottom of the battery compartment.
4. Return the battery door to the unit and tighten the screw to secure the battery
door.
Do not over tighten the battery door. Doing so can damage the test unit.
Cleaning
Use a damp, clean cloth to clean the test unit.
Disconnect all cables from the FaultChaser prior to cleaning. Failing to do so can
damage the unit and result in personal injury.
Do not use abrasive, harsh cleaners, or solvents to clean the FaultChaser.
Storage
When the
FaultChaser
is not in use, store in a dry, protective case. The batteries should
be removed if the unit is stored for a long period of time.
Do not expose the
FaultChaser
to high temperatures or humidity. When stored in
temperatures exceeding the limits listed in the
Specifications
section, allow the
FaultChaser
to return to the normal, recommended operating conditions prior to use.
27