26-0008-1100/5
(Worldwide)
— 29 of 92
Physician’s Manual — VNS Therapy Programming Software (8.1)
Figure 17.
DEVICE DIAGNOSTICS Menu
Use this screen to select function assessments to be performed on the Pulse
Generator. The System Diagnostics, as well as the DC-DC Converter
Code, can be used at patient follow-up visits to assess lead impedance
status for Models 100, 101, and 102. An actual impedance measurement is
reported for Models 103, 104, and 105. The System Diagnostics is also
used during implantation and routine follow-up visits to assess nerve-
electrode impedance. The Normal Mode Diagnostics, Magnet Mode, and
Generator Diagnostics can be accessed by tapping
Other Diagnostics
(see