Physician’s Manual — VNS Therapy Programming Software (8.1)
and its effect on device parameters. The parameters marked with asterisks
indicate unintentionally altered or incorrectly displayed values.
Table 3.
Cross-programming Example
*Unintentionally altered or incorrectly displayed values.
Due to the susceptibility of Model 100, 101, and 102 Pulse Generators to
cross-programming, LivaNova recommends that an initial and a final
interrogation be performed at each office visit for all VNS patients in order
to verify parameter settings.
4.2.
PARAMETER Screen Menu
Tapping the
Menu
button at the right top of the PARAMETER Screen
causes the PARAMETER Screen Menu to be displayed, as shown in
Figure 16 on page 28.
Parameter
After Step 2
At Step 4
After Step 4
Patient A Final
Settings
Patient B Initial
Settings
Patient B
Intended Final
Settings
Patient B
Actual Final
Settings
Patient A Final
Settings in
Database
Output Current
(mA)
0.25
0.75
No Change
0.25*
0.25
Frequency
(Hz)
30
20
15
15
15*
Pulse Width
(µs)
250
500
No Change
500
250
On Time (s)
30
7
No change
7
30
Off Time (min)
5
1.8
3
3
3*
Mag. Output
Current (mA)
0.5
1.00
No Change
0.5*
0.5
Mag. On Time
(s)
30
30
60
60
60*
Mag. Pulse
Width (µs)
500
500
250
250
250*