lion alcolmeter® 900
User Handbook Ref: RM92905 Issue: 1.0
7
Technical Details
Factor
Target
Comments
Specificity
Specific to alcohol
Not sensitive to
acetone, paint and
glue vapour, CO, food,
sweets, methane, and
non- alcoholic
materials which may
be found in human
breath
Measurement
range
0.000 %BAC
Note 2
Minimum
Note 1
0.500 %BAC
Note 2
Maximum
Measurement
Accuracy
<±0.005 %BAC or <±5%
(whichever is greater)
In Active Mode
Measurement
step
0.001 %BAC
0.0001 %BAC
(metrological mode)
Response
time (zero)
≤
2 seconds
at 20
o
C and above
Response
time
< 10 seconds
Recovery
time
< 30 seconds
at 0.080 %BAC
Sampling
frequency
< 10 seconds
Zero test result