Praktika 2
Positionable hospital bed

D9U001F20-0101

Version: 07

Publication Date: 2020-02

Instructions for Use 

and Technical Description

Summary of Contents for Praktika 2

Page 1: ...Praktika 2 Positionable hospital bed D9U001F20 0101 Version 07 Publication Date 2020 02 Instructions for Use and Technical Description...

Page 2: ...0101 Version 07 Publication Date 2020 02 Copyright LINET s r o 2020 Translation LINET 2020 All rights reserved All trademarks and brands are the property of the appropriate owners The manufacturer re...

Page 3: ...tioning of calfrest 25 10 3 5Trendelenburg Antitrendelenburg position 26 10 4Quick Backrest Lowering for the Crisis Situations 27 11Siderails 28 11 1Single Collapsible Siderails Controlled From Bottom...

Page 4: ...arns about the risk of material damage WARNING warns about the risk of physical injury DANGER warns about the risk of fatal injury 1 1 2Structure of Warning Notices SIGNAL WORDS Type and source of dan...

Page 5: ...D9U001F20 0101_07 5 1 4Symbols on the Package FRAGILE HANDLE WITH CARE THIS WAY UP KEEP DRY PROTECT FROM HUMIDITY PAPER RECYCLING SYMBOL DO NOT USE HAND TRUCK HERE DO NOT STACK DURING STORAGE...

Page 6: ...OR USE TO INFORM YOURSELF ABOUT RIS KS WHEN REMOVING HEAD BOARD OR FOOT BOARD DESCRIPTION HOW TO CONTROL LEVER MECHANISMS TRENDELENBURG Antitrendelenburg POSITION DO NOT PLACE ANY OBJECTS UNDER MATTRE...

Page 7: ...ONFORMITY WITH EU REGULATION DESIGNATION OF HOSPITAL BED FOR ADULTS MEDICAL DEVICE COMPATIBLE WITH MEDICAL DEVICE REGULATION WEEE SYMBOL RECYCLE AS ELECTRONIC WASTE DO NOT PUT INTO THE HOUSEHOLD WASTE...

Page 8: ...D9U001F20 0101_07 8 SERIAL NUMBER...

Page 9: ...Day 3 DI Device Identifier GTIN Global Trade Item Number 4 1D Bar code GS1 128 Serial Number 5 Symbols 6 Configuration number 7 Electrical Specification 8 Serial Number 9 PI Product Identifier 10 2D B...

Page 10: ...accessories and the load supported by those accessories Siderail Height the height of the upper crossbar or the edges of the siderails not the highest point of the siderail controls from the patient s...

Page 11: ...ts and bed lifts WARNING To avoid risk of electric shock this equipment must only be connected to a supply mains with protective earth WARNING During specific investigations or treatments the signific...

Page 12: ...is not intended for oxygen enriched environment WARNING This medical device is not intended for use with flammable substances WARNING This medical device is not portable medical electrical equipment...

Page 13: ...of whether the bed is occupied or empty Ensure that siderails are operated by healthcare personnel only Never use the bed in areas where there is a hazard or risk of explosion Ensure moving parts of...

Page 14: ...ses doctors technical personnel transport personnel cleaning personnel 3 2Contraindications The medical device is not intended for the pediatric patients use Certain positions are not suitable for spe...

Page 15: ...ic siderail in highest positions 4 Head board 5 Head siderail release mechanism 6 Protective bumper 7 Castor 150mm 8 Center siderail release mechanism 9 Accessory rail 10 Cental brake system lever 11...

Page 16: ...ngle collapsible siderails with Extender 15 cm 16 cm 26 cm Space underneath the bed Mattress platform down 8 5 cm 15 cm since mattress platform height 49 5 cm Mattress platform height 125 mm castors 1...

Page 17: ...the serial label s standard o optional Mattress platform Mattress platform consisting of steel parts s Mattress platform consisting of removable plastic parts o Siderails Without siderails s Split pla...

Page 18: ...utting out of service and maintenance CAUTION Material damage due to incorrect putting into service Ensure that putting into service is performed exclusively by manufacturer s customer service or trai...

Page 19: ...and there are no accessories atta ched to the head board and foot board If a patient is lying in a bed with the head board and or foot board removed supervise the bed at all times WARNING Material dam...

Page 20: ...ible to remove head board and foot board Check the functionality of siderails 9 4Transport Observe the following for safe transport Ensure the castors are unlocked before moving the bed during the loa...

Page 21: ...Antitrendelenburg positioning Fig Control levers 10 1Positions of Levers WARNING Risk of damaging the bed during transport Always set both levers into transport position before transporting the bed o...

Page 22: ...n moving parts of the mattress platform and undercarriage frame Secure or remove any items on the bed Ensure no persons or body parts are close to the bed or accessories e g infusion stand lifting pol...

Page 23: ...backrest Fig Lowering the backrest 10 3 2Thighrest Positiong It is necessary to set lever into its working position see Manipu lation with levers Fig Thighrest positioning Fig Raising the thighrest R...

Page 24: ...set lever into its working position see Mani pulation with levers Fig Thighrest positioning Fig Raising the bed height Raise the bed height as follows Set the left lever to its working position By ro...

Page 25: ...ed mechanically using raster 1 Mattress holder Calfrest handle 2 Mechanical raster Raise the calfrest as follows Grab the calfrest handle 1 Pull the handle upwards until desired position is reached En...

Page 26: ...e Trendelenburg position as follows Ensure head board and foot board are locked Position the bed to approx 20 cm above the lowest position Pull and hold the red lever towards yourself and simultaneous...

Page 27: ...osition Ensure there are no body parts between siderails and backrest The bed allows quick mechanical lowering of the backrest using quick counterclockwise rotatition of the right lever Backrest will...

Page 28: ...ent falling out of bed Ensure that folded up siderails are securely anchored Press against siderails from inside in order to check stability Hospital personnel are responsible for ensuring siderails a...

Page 29: ...k when locked in place Ensure that the siderail is locked properly 12Castor Control and Bed Transport WARNING Material damage due to incorrect transport or involuntary movement Prior to assembly disas...

Page 30: ...ory rail Fig Castor with central brake system The bed is equipped with central brake system Braking levers are lo cated on the both sides of the bed Central brake system is designed for easier castor...

Page 31: ...MACARE 20 CLINICARE 10 CLINICARE 20 CLINICARE 30 15Accessories WARNING Risk of injury due to incompatible accessories Only the manufacturers original accessories can be used The manufacturer assumes n...

Page 32: ...backrest 4 working position Parallel to the head board when not in use NOTE Lifting pole is not a part of delivery and needs to be ordered separately NOTE The date of manufacture is marked on the pla...

Page 33: ...as driving pushing device during the bed transport In case of failure to comply those instructions there is a risk of collision of accessories with movable parts of the bed or injury of the patient 1...

Page 34: ...ot board so the Protector is facing out from the bed Uninstall Protector as follows Press and hold the releasing button 1 Pull the Protector upwards 1 Closed position 2 Opened position Fig Protector p...

Page 35: ...is adapte for siderails It is recommended to use Extender when placing higher mattress on the bed than recommended by manufacturer Extender will raise the height of siderails and also safety of the pa...

Page 36: ...are compatible with materials that the product consists of For information see the following table BED COMPONENTS THAT ARE INTENDED TO BE CLEANED Do not clean what is not mentioned in this column MAT...

Page 37: ...efore changing patient or complete Let the bed dry completely after cleaning or disinfection process After drying the bed place the mattress back on the mattress platform After drying the bed check fu...

Page 38: ...t and its dilution with manufacturer of the detergents according to the material table below Part of bed Praktika 2 Daily C D Changing patient C D Complete C D Siderails stickers Split plastic Single...

Page 39: ...l is not possible The siderail lock is dirty Clean the locking mechanism Defective siderail lock Notify service department Faulty brakes The brakes are blocked by dirt Clean the brake system Defective...

Page 40: ...ts by the original spare parts Check regularly that all accessories are working properly Replace damaged accessories immediately 18 2Spare Parts The serial label is located on the frame of the mattres...

Page 41: ...s of noise and vibration in protected interior spaces of buildings according to stan dard IEC 60601 2 52 Used packaging materials are in accordance with requirements of the Packaging Act Z kon o obale...

Page 42: ...l material and manufacturing re lated failures and errors Failures and errors caused by incorrect use and external effects are not covered Justified complaints will be fixed free of charge during the...