52
[Adjustments]
①
If the bubble of the plate lever moves from the center, bring it half way back to the
center by adjusting the leveling screw(s) which is parallel to the plate level.
②
Correct the remaining half by adjusting the bubble adjusting nuts with the
adjusting pin.
③
Confirm that the bubble does not move from the center when the instrument is
rotated by 180°.
④
When the bubble moves, repeat from (1) once again..
11.2 Circular vial
[Checks]
①
Adjust by the plate level vial beforehand.
②
Confirm the position of the bubble of the circular vial.
At this time, it is not necessary to adjust if the bubble is at the centre of the circle.
[Adjustments]
When the bubble of the circular vial comes off from the centre according to check procedure
②
, it is necessary to adjust.
Leveling screw to be adjusted.
Plate vial
Leveling screw to be adjusted.
Plate level
Bubble adjusting screw