6.7 Auto Run Function
By using this function, you can repeat the measurement automatically for the specified period by specifying the date
and time when the measurement starts and ends. It can start the measurement in conjunction with power-on and auto
-
matically finish the measurement at the specified date and time.
Setting
Display the top menu screen by [MENU], touch “Record control
”, and then touch the “Auto run” tab to display
the setting screen.
ڦ
Mode
Select the measurement repetition period from Monthly (run
monthly), Daily (run daily), or Hourly (run hourly).
ڦ
Run time
Set the date and time when to start the measurement accord-
ing to the mode. It becomes effective when checked.
ڦ
Stop time
Set the date and time when to finish the measurement according to the mode. It becomes effective when
checked.
You can use it by checking one or both of “Run time” or “Stop time”.
ڦ
External power on Run
When checked, measurement starts 10 seconds after the start of USB bus power supply. It is used when you
want to start measurement in conjunction with the turning on by the external power.
Control
When “Run time” is checked, after pressing [RUN] it turns to be
wait status until the specified time arrives. To cancel the waiting
status press [STOP] or touch “Cancel”.
When the specified time comes, the measurement starts auto-
matically. When the “Stop time” is checked, the measurement
is performed up to the specified time and the measurement au
-
tomatically stops. This measurement process will be executed
repeatedly according to the condition set at “Mode”.
When “External power on Run” is checked, when USB bus power
supply is started while the power of this unit is turned off, the
power is automatically turned on, and 10 seconds after that, the
measurement starts automatically without pressing [RUN] .
This function does not work when USB bus power is continu-
ously supplied.