CLEAR SHIELD HOOD PAPR UNIT

GENERAL CLEANING GUIDELINES 

K4886-1, K4886-3

Maintenance and Storage

•  Read and follow all steps set forth in the 

Clear Shield Hood PAPR Unit Operator’s 

Manual.

•  Replace damaged or dirty air filters. 

Filters cannot be washed or cleaned with 

compressed air. Never reuse a dirty air filter.

•  Never use solvents or abrasive cleaning 

solutions to clean the respirator. Keep water 

and other fluids out of blower assembly.

•  Maintain accurate records of filter 

replacement and respirator maintenance.

•  Cleaning is recommended after each use. 
•  After the whole set of the product is 

completely dry, connect the hood with 

the blower unit, install the new filters or 

accessories, and store it in a disposable 

medical plastic bag.

IMPORTANT NOTICE:

•  Follow your facility’s protocols for proper 

disposal of cleaning equipment and solutions.

Recommended Disinfection Method

Sodium hypochlorite disinfection method 

(free chlorine concentration is 5000 ppm 

(0.5%)). This concentration standard is 

derived from the “Guidelines for Temporary 

Infection Control for Nursing Suspected or 

Diagnosed Filamentous Hemorrhagic Fever 

in Medical Institutions” issued by the WHO 

(World Health Organization) in August 2014. 

http://www.who.int/csr/resources/who-ipc-

guidance-ebolafinal-09082014.pdf?ua=1

Product Disinfection Handover

The Clear Shield Hood PAPR Hood can be 

handed over using the following steps:

1. 

Prepare a graduated spray bottle (1.0- 1.5L 

volume) with 1L clean water inside. Making 

the disinfecting solution of 5000mg/L by 

adding 10pcs of LIRCON effervescent tablets 

with an effective chlorine content of 500mg/

tablet and mix thoroughly. (For reference 

only, subject to actual conditions of each 

hospital). 

2. 

When medical workers wearing the Clear 

Shield Hood PAPR leave isolation section, 

please make sure the blower unit is always 

working (

DO NOT

 turn it off). Other medical 

workers spray the whole hood PAPR system 

(hood, hose, belt, and blower unit) with 

chlorine-containing disinfectant.

3. 

After spraying disinfectant, turn off the 

blower unit, take off the hood (from inside 

out), blower unit, and hose and pack them 

with the clean medical plastic bags. Put them 

in the designated area of the buffer zone, 

and take them to the disinfection area by a 

designated person for disinfection.

Disinfection Process in  

Designated Area

Preparations:

1. 

Disinfectors must wear a full set of 

protective equipment including protective 

clothing, mask, and goggles.

2. 

Prepare two layers of nitrile gloves or vinyl 

gloves before cleaning.

3. 

Prepare 5000mg/L chlorine-containing 

disinfectant and medical gauze.
Initial Process:

1. 

Get the complete protective equipment out 

from the disposable medical bag.

2. 

Separate each part (hood, hose, blower unit) 

carefully.
Hood Disinfection:
The Clear Shield Hood PAPR Hood can be 

cleaned following the steps below:

1. 

Remove the sweatband from the headband 

inside the hood, put the sweatband into a 

disposable medical plastic bag and discard.

2. 

Clean the hood inside and outside surface 

three times by using the medical gauze 

soaked with chlorine disinfectant (0.5%). 

Dip the neutral (PH6-8) cleaner to wipe the 

hood and use the clean water to wash off the 

remaining liquid of the cleaner.

3. 

Put the disinfected hood into a disposable 

medical plastic bag and dry them in a 

pollution-free area. Please make sure to 

assemble the new sweatband correctly 

before next use.

IMPORTANT NOTICE:

•  If there is sufficient inventory for the hood, 

one-time use is recommended.

•  The cleaned Hood should be carefully 

inspected per the Operator’s Manual.

•  Store Helmet when not in use in accordance 

with the Operator’s Manual.

Blower Unit Disinfection

The Clear Shield Hood PAPR Blower Unit can 

be cleaned by following the steps below:

1. 

Clean the surface of the blower unit 

(including the belt) and the hose connection 

area three times by using the medical gauze 

soaked with chlorine disinfectant (0.5%).

a. 

Note: When the complete hood PAPR 

system is running, the air is always clean 

after filtration, so only need to clean the 

hose connection area which is exposed to 

the air.

2. 

Please clean the white area inside the blower 

unit carefully. 

a. 

DO NOT

 wash it with liquid, the liquid may 

penetrate into the equipment and cause 

damage.

CONTINUED ON BACK >>>