9.20. Motor voltage [107]
Reading this command will return the voltage drawn by the motor in mV.
Read
Write
Persists
ui nt 16
-
-
9.21. Output rate [108]
The outputrate controls the amount of data sent to the host when distance outputstreaming is enabled.
Value
Points per
second
0
20010
1
10005
2
6670
3
2001
Read
Write
Persists
ui nt 8
ui nt 8
Yes
9.22. Forward offset [109]
The forward offset affects the positionof the
0 degr ee
direction. The
or i ent at i on
label on the frontof the
SF40/C marks the default
0 degr ee
direction.
Read
Write
Persists
i nt 16
i nt 16
Yes
9.23. Revolutions [110]
Reading this command will return the number of full revolutions since start-up.
Please note that this value will reset to zero after 4294967295revolutions.
Read
Write
Persists
ui nt 32
-
-
SF40/C
Product manual
SF40/C -Product manual
|
26 of 30 |
© LightWare Optoelectronics (Pty) Ltd, 2019 |
www.lightware.co.za