M70C   
Operator’s Manual Fingertip Pulse Oximeter 

Version number of this manual: V1.0                    Document No. : J/M70-O-346-001

 

   

General Description                                                                         

The  measurement  of  oxygen  saturation  of  arterial  blood  (also  known  as  pulse 
oxygen  saturation,  usually  shortened  as  SpO

2

)  adopts  the  principles  of  light 

spectra and volume tracing. The LED emits lights with two specific wavelengths, 
which are selectively absorbed by oxygenated hemoglobin and deoxyhemoglobin. 
The  optical  receptor  measures  the  changes  in  the  light  intensity  after  the  light 
passes  the  capillary  network  and  estimates  the  ratio  of  oxygenated  hemoglobin 
and the total hemoglobin. 

 

 

 

Method  for  evaluating  SpO2  accuracy:  The  SpO2  accuracy  of  pulse  oximeter  is 
measured by comparing SpO2 readings of the pulse oximeter to values of SaO2 
determined with a CO-OXIMETER.The healthy volunteers who consent to induced 
hypoxia  and  arterial  blood  sampling  as  part  of  the  experimental  procedure  (see 
ISO80601-2-61 Annex EE.2) 

    Caution                                                                                           

  Please read the user manual carefully prior to operating. 

   

Intended Use                                                                                             

The  Fingertip  Pulse  Oximeter  is  intended  to  measure  functional  arterial  oxygen 
saturation  (SpO2)  and  pulse  rate  of  adult,  pediatric  and  adolescent  patients  in 
hospital, hospital type facilities as well as in the home care environment. 
The oximeter is not suitable to monitor patient continuously for long term 

    Battery Installations                                                                             

1.

 Push the battery cover. 

2.

 Install two AAA batteries into battery cabin in correct 

polarities, as shown on the right. 

3.

 Close the battery cover. 

Notes: 

 

For battery installation, see picture to the right. 
Put or remove batteries in right order, or it may 
damage the bracket.   

 

Battery polarities must be correctly installed. Otherwise, damage might be 
caused to device. 

 

Please remove the battery if the oximeter will not be used for long time. 

   

Operation Instructions                                                                             

1.

 Install two AAA batteries into battery cassette before closing its cover. 

2.

 Nip  the  oximeter,  then  insert  one  of  fingers  into  the 

rubber  hole  of  the  oximeter  before  releasing  the 
oximeter, and your nail surface must be upward. 

 

3.

 Press the function button once on front panel. 

 

4.

 Your finger and body should not tremble during measuring.   

5.

 Read corresponding data on the display screen. 

6.

 After  turning  the  oximeter  on,  each  time  you  press  the  power  switch,  the                   

display screen will change to another direction. There are two display modes. If you 
long press the power switch, you can adjust the brightness of screen from 1 to 5. 

 

 
 
 
 
 

 

SpO

2

 Plethysmogram 

 

SpO

reading 

 

Pulse rate reading 

 

Indication of battery capacity 

 

Indication of pulse intensity 

 

Indication of screen brightness 

 

Display modes 

   

Precautions for Use                                                                           

  The patient is the operator when the device is used at home. 

  Patients can maintain and use all functions of the device safely according to this 

user′s manual. 

  Keep this product out of reach of children to avoid injury to children. 

  Explosion hazard. Do not use the oximeter in the presence of flammable 

anesthetics mixture with air, oxygen, or hydrogen. 

  When the oximeter is in use, there should not be any great power appliances as 

high voltage cables, X-ray machine, ultrasound equipment and electrizer in use 
nearby.   

  Keep  the  oximeter  away  from  lint,  dust,  vibration,  corrosive  substances, 

explosive materials, high temperature and moisture. 

  This oximeter does not have an alarm function; please do not use this product in 

the environment where an alarm is required. 

  The oximeter should be handled with care so as to avoid shocks and falls.   

  When  the  oximeter  is  in use,  it  must  be  ensured  the batteries have  sufficient 

capacity;  otherwise,  there  might  be  such  phenomena  as  starting-up 
abnormalities or inaccurate measurement data, etc.     

  Please  do  not  use  pointed  objects  such  as  pen  point  or  nails  for  pressing 

operation,  otherwise  it  might  cause  permanent  damage  to  the  surface  of  the 
keyboard. 

  Do not make any clinical judgments based solely on the oximeter. The oximeter 

is intended only as an adjunct in patient assessment. It must be used in 
conjunction with clinical signs and symptoms, as well as doctor’s diagnoses. 

  To ensure accurate performance and prevent device failure, do not expose the 

oximeter to extreme moisture, such as direct exposure to rain. Such exposure 
may cause inaccurate performance or device failure. 

  Do  not  conduct  SpO

2

  measurement  on  the  finger  smeared  with  nail  polish, 

otherwise this will lead to unreliable measurement results. 

  Please do not open the enclosure. The enclosure shall only be opened by the 

authorized person. 

  In order to have more accurate measurements of SpO

2

 and PR, the oximeter 

should be used in quiet and comfortable environment. 

  Follow  local  ordinances  and  recycling  instructions  regarding  disposal  or 

recycling of the device and device components, including batteries. 

  Prolonged continuous monitoring may increase the risk of unexpected changes 

in skin characteristics, such as irritation, redness, blisters or burns. Inspect the 
sensor site every two hours and move the sensor if the skin quality changes. 

  Pulse oximeter simulator can not be used to access the accuracy of the pulse 

oximeter. 

  The expected service life of the device is five years. 

  For assistance with installation, use or maintenance, contact the manufacturer 

or manufacturer's representative. 

  To validate the PR accuracy, we refer to the electronic pulse simulator. 

  When used at room temperature from the lowest or highest storage temperature 

environment, the product can be used directly without needing to be placed for 
a period of time to achieve its expected function. 

    Download and Install APP Software                                                                     

1.

  Software download methods

 

  IOS device users, login to your 

App Store

 account and search for LifeSmart 

Health.   

  Android  device  users,  login  to  your 

Google Play

  account  and  search 

LifeSmart.   

QR Code for download:   

                          Android system                                        IOS system 
 
 
 

 

 

 

                               

 

2.

  Download the 

Free LifeSmart App

 

Note: App software only applied for the oximeter with Bluetooth. 
 

   

Maintenance                                                                                         

1.

  Using a soft cloth dampened with either a commercial nonabrasive cleaner, or 

a solution of 70% alcohol in water, lightly wipe the surfaces of the oximeter. 

2.

  The  most  commonly  used  hospital  cleaning  agent  and  non-corrosive 

detergents can be used for cleaning the oximeter, but please be careful that 
most detergents must be diluted before use; Please use them according to the 
directions of the manufacturers of the detergents.   

3.

  Avoid using alcohol-based, amido or acetone-based detergents   

4.

  The casing of the oximeter should be kept from the contamination of filth and 

dirt, and it can be wiped with non-velvet soft cloth. When cleaning, do not spill 
the liquid onto the instrument. Ensure no liquid is allowed to enter the inside of 
the oximeter.     

5.

  It is forbidden to use such grinding materials as wire brush or metal polishing 

agent,  because  these  materials  may  cause  damage  to  the  panels  of  the 
oximeter.   

6.

  Please do not soak the oximeter in liquid. 

7.

  Under  normal  circumstances,  it  is  unnecessary  for  the  oximeter  to  have 

special maintenance, and cautions must be exercised on the following points 
during the use of the oximeter:   

  Please  use  the  oximeter  in the environment according to the requirements of 

the performance criteria. 

  Avoid exposure to direct sunlight. 

  Avoid excessive radioactive infrared rays or ultraviolet rays. 

  Avoid contacts with organic solutions, dusts or corrosive gases. 

 

   

Product Specifications                                                                       

 

Measurement specifications 

SpO

2

 

Measuring Range 

0~100% 

Resolution 

1% 

Accuracy 

At 70~100%, ±2%

 

At 0~69%, unspecified   

Data update period 

<13 s 

%

100

lobin

deoxyhemog

 

bin 

oxyhemoglo

hemoglobin

 

oxygenated

%

SpO

2

 

 

 

 

 