Process Fluid Chiller Designed For Varian CLINAC 4/100, 6/100 & 600C 2
1050 Dearborn Drive
P.O. Box 29186
Columbus, OH 43229
Phone: 800.877.9222
Fax: 614.841.6022
Via Leonardo Da Vinci 8
Zona Industriale Tognana
35028 Piove Di Sacco
Italy
Phone: 39.049.9719.111
Fax: 39.049.5841257
23/F Allied Kajima Bldg.
138 Gloucester Road
Wanchai
Hong Kong
Phone: 852.2.572.2201
Fax: 852.2.572.1182
©2000 Liebert Corporation. All rights
reserved throughout the world.
Specifications subject to change without
notice.
®Liebert and the Liebert logo are
registered trademarks of Liebert
Corporation. All names referred to are
trademarks or regisered trademarks of
their respective owners.
Model No.
A-Depth
in. (mm)
B-Width
in. (mm)
C-Height
in. (mm)
Weight
lbs. (kg)
Voltage
(60 Hz)
Voltage
(50 Hz)
PS018A
36 (914)
40 (1016)
25.25 (641))
394 (178)
208/230, 1-phase
200/230, 1-phase
PS024A
36 (914)
40 (1016)
25.25 (641)
404 (183)
208/230, 1-phase
200/230, 1-phase
PS036A
36 (914)
48 (1219)
32.5 (825)
515 (233)
460, 3-phase
208/230, 1- or 3-phase
400, 3-phase
200/230, 1- or 3-phase
PS048A
36 (914)
53 (1346)
38.5 (978)
592 (268)
460, 3-phase
208/230, 3-phase
400, 3-phase
200/230, 1- or 3-phase
PS060A
36 (914)
53 (1346)
38.5 (978)
642 (291)
460, 3-phase
208/230, 3-phase
400, 3-phase
200/230, 1- or 3-phase
ACCESS, AND AIR FLOW.
FOR COMPONENT REMOVAL,
RECOMMENDED CLEARANCE
INDICATE A
SHADED AREAS
H
H
I
I
(25mm)
1" MAX.
(25mm)
1" MAX.
(914mm)
36"
(914mm)
36"
UNIT DIMENSIONAL DATA
OUT
AIRFLOW
B
IN
AIRFLOW
C*
D
A
1" NPT
(914mm)
36"
(51mm)
2"
(457mm)
18"
ACCESS, AND AIR FLOW.
FOR COMPONENT REMOVAL,
RECOMMENDED CLEARANCE
INDICATE A
SHADED AREAS
PRESSURE GAUGES
OPTIONAL REFRIGERANT
UNIT DIMENSIONAL DATA
Standard Features
• Indoor/outdoor cabinet
• Hermetic compressor
• Refrigerant dehydrator
• Expansion valve
• Refrigerant receiver
• Stainless steel fluid reservoir expansion tank
• Pressure relief valve
• Hot gas bypass valve capacity control
Optional Features
• Refrigerant gauges and reservoir heater
• Stainless steel pump
Accessories
• Emergency water bypass
• Enviro-source emergency bypass
• Water level control and alarm system
• Fluid pressure gauge
• Check valve
• Dial thermometer
• Relief valve
• Circuit setter
• Ball valve
• High temperature thermostat
• Low flow switch
• Liebert RCM8 monitor
Liebert Web Site—http://www.liebert.com
SL-16548 (07/00)