Section 3. Operation
If you log all variables at 10 Hz (10 samples per second), approximately 180 MB of
data will be generated each day. Thus, the 16 GB drive can collect about 80 days of
data with no compression. With compression, a daily file would be about 50 MB;
the 16 GB drive can collect about 288 days of data.
Transferring Logged Data
Files can be using the Download utility that is built into the gas analyzer PC soft-
ware or a File Transfer Protocol (FTP) application.
The Download utility allows you download and delete files that are stored on the
USB drive. You can access the files remotely, so long as you can connect to your
instrument over a network. The file transfer program is a component of the gas ana-
lyzer software. Click the
Download
button to launch it.
The application should automatically connect to the instrument, but if it doesn't,
click
Browse
to view a list of instruments on the network. Select the instrument,
click
OK
, enter the password (
licor
) and click
Verify Connection
. You may need to
3-46
LI-7200RS Enclosed CO₂/H₂O Analyzer
Summary of Contents for LI7200RS
Page 1: ...Enclosed CO2 H2O Gas Analyzer Instruction Manual...
Page 2: ......
Page 11: ...ix Appendix C Suppliers Appendix D Configuration Grammar Appendix E Warranty Appendix F Index...
Page 12: ...x...
Page 20: ...Section 1 General Information 1 8 LI 7200RS Enclosed CO H O Analyzer...
Page 54: ...Section 3 Operation 3 12 LI 7200RS Enclosed CO H O Analyzer...
Page 112: ...Section 4 User Calibration 4 12 LI 7200RS Enclosed CO H O Analyzer...
Page 130: ...Section 5 Maintenance 5 18 LI 7200RS Enclosed CO H O Analyzer...
Page 138: ...Section 6 Troubleshooting 6 8 LI 7200RS Enclosed CO H O Analyzer...
Page 184: ...Section 8 Theory of Operation 8 12 LI 7200RS Enclosed CO H O Analyzer...
Page 187: ...Gas Analyzer Dimensions Appendix A Specifications A 3 Appendix A Specifications...
Page 188: ...Appendix A Specifications LI 7550 Dimensions A 4 LI 7200RS Enclosed CO H O Analyzer...
Page 192: ...Appendix B Pin Assignments B 2 LI 7200RS Enclosed CO H O Analyzer...
Page 196: ...Appendix C Suppliers C 4 LI 7200RS Enclosed CO H O Analyzer...
Page 225: ......